µPG 501
34
User Guide
Remedies:
- check design
- check wizard entries
Problem:
Design looks completely mixed up
Possible causes:
- vacuum off
Remedies:
- check vacuum pump and vacuum on stage
Problem:
Design stretched or shrunk compared to previous layer
Possible causes:
- previous layer not exposed with µPG, coordinate system units do not match
Remedies:
- shrink or stretch design to compensate
ADVANCED TROUBLESHOOTING
When informing a service office about a problem with the µPG 501 to get further
troubleshooting advice, please include the following:
•
Information on settings of compressed air regulators
•
Any error numbers and error descriptions provided by the wizard
•
The wizard version that is shown when selecting
About
in the menu bar
•
If available, screenshots / pictures of the problem, or descriptive sketches
•
If the problems is connected to an exposure, create an extended report file by
activating the option create detailed report file
•
Send the folder
C:\HIMT
in archived form (.zip, .rar) to your service office. If the file
gets too large or archiving takes too long, leave out the subfolder
LICSources
.