MLA150
6 Job Setup
58
User Guide
When choosing
Dose Series
only, the Defoc value is fixed. The Dose value is fixed
when
Defoc Series
is active.
The substrate visualization shows the arrangement of the fields according to the values
set in the
Series
panel. This feature facilitates the preparation of series exposures. The
horizontal fields are for defocus series and the vertical fields represent the dose fields
(see below). If the fields overlap, the step size has to be changed to a higher value. If
the field arrangement exceeds the substrate surface size, the number of fields has to be
reduced.
Select from the options in the
Mode
dropdown list:
Dose
,
Defoc
or
Dose
and
Defoc
.
Set the values required for the selected mode.
Note:
Pay attention to the value in the Total range fields. If the total range
requires more space than available on the substrate, an error message
pops up.
If desired, activate
Auto-Unload
(see
6.9 Unload Substrate
).
Click
Start Exposure
.
Содержание MLA150
Страница 1: ...User Guide MLA150 Maskless Aligner...
Страница 2: ...MLA150 ii User Guide Doc No DWL HI 062 Revision 1 July 2016 Copyright 2016 by Heidelberg Instruments...
Страница 10: ...MLA150 1 Introduction 4 User Guide...
Страница 16: ...MLA150 2 Safety 10 User Guide 2 7 Labels Flowbox front E Rack REAR DOOR FRONT DOOR...
Страница 17: ...MLA150 2 Safety User Guide 11 Optics cover front cover side covers rear cover...
Страница 18: ...MLA150 2 Safety 12 User Guide...
Страница 68: ...MLA150 6 Job Setup 62 User Guide...
Страница 75: ...MLA150 7 System Utilities and Tools User Guide 69 Figure 16 BSA marker zones overview...
Страница 80: ......
Страница 81: ......