RMLD-CS
Operator’s
Manual
Doc No. 105408 Rev C, RMLD-CS Operator Manual
© 2021 Heath Consultants Incorporated. All rights
reserved.
9
Specifications
Detection Method
Tunable Diode Laser Absorption Spectroscopy (TDLAS)
Measurement Range
0 to 50K PPM-M. Concentration units shift at 10000 PPM-M and
higher to K PPM-M; example 9999, 10K, 12K, 20K, 50K
Sensitivity
5 PPM-M at distances from 0 to 100 ft (30m)
Detection Distance
100 ft (30m) nominal. Actual distance may vary due to
background type and conditions.
Beam Size
Conical in shape with a 22” diameter at 100 ft (55cm at 30m)
Detection Alarm Modes
Digital Methane Detection (DMD): Audible and visual alerts when
the alarm level is exceeded. Adjustable Alarm Level from 1 to
999 PPM-M.
Real Time (RT): Continuous audio beep rate relative to
concentration from 10
–
1000 PPM-M.
System Fault / Warning
Audible alarm and visual indication on the display.
Self-Test
Built-in Self-Test function verifies operation and adjusts laser
wavelength for maximum sensitivity. Cal gas cell integrated
within carrying case.
Compliance
EMC (EN61000-6-2, EN6100-6-4)
Radio Equipment Directive (2014/53/EU)
ETSI EN 301 489-1 v2.2.0
EN 61326-1:2013
47 CFR Part 15 & ICES-003
Laser Eye Safety
IR Laser: Class I
Visible Laser: Class II
Do not stare into beam or view directly with optical instrument
Communications
USB Dual Mode
–
flash drive and communications
Bluetooth 4.2 BLE
Wi-Fi b/g
Display
3.5" LCD Color
320 x 240
Operating Temperature
0° to +122° F (-17° to 50° C)
Humidity
5 to 95 % RH, non-condensing
Instrument Protection
IP54 (water splash and dust resistant)
Содержание HPN105354