Doc No. 105408 Rev C, RMLD-CS Operator Manual
RMLD-CS
Operator’s Manual
Manual
© 2021 Heath Consultants Incorporated. All rights
reserved.
62
Excessive false detection
or loss of sensitivity
Laser output not optimized
Perform Self-Test procedure
Alarm Detection Threshold set too
Check the Alarm Detection
high or low for conditions
Threshold
Error message or
Warning icon on
continuously
Low battery
Check battery level and recharge if
necessary
Moisture condensation on mirror due to
rapid change in temperature Internal
component failure
Allow for the temperature of
the instrument to stabilize
Internal component failure
Note error message and contact HEATH
Reduced run time
Battery not fully charged
Charge battery pack until solid green
light on charger is on
Diminished battery capacity
Replace battery pack, and properly
dispose of exhausted battery pack.
Low Signal or Low Light
Background surface is absorbing or
reflecting the IR light level
Change angle to the target to get a
better reflecting background
Scanning at the range limit of the
instrument
Move in closer
Saturated
Background surface is reflecting the IR
light level
Change angle to the target to get a
better background
No Wi-Fi connection Incorrect credentials
Verify credentials
Outside of Wi-Fi range
Place unit closer to Wi-Fi point
Wi-Fi disabled in menu
Enable Wi-Fi in menu
No Bluetooth connection Bluetooth is disabled
Enable Bluetooth in menu
Device not paired
Pair device with instrument
USB not recognized USB cable does not work
Use another USB cable
Port is not working
Try a different USB port or allow
computer to load drivers
Low battery
Charge or replace battery pack
Driver did not install properly
Uninstall device driver and allow to
reinstall
IT restrictions
Contact local IT support