Doc No. 105408 Rev C, RMLD-CS Operator Manual
RMLD-CS
Operator’s Manual
Manual
© 2021 Heath Consultants Incorporated. All rights
reserved.
43
When edits are
done, click the “
Save Users to RMLD
” button
on the bottom row
or the changes won’t be
saved. A confirmation message will pop up.
NOTE:
Should the created ADMIN password be lost or forgotten, please contact Heath Customer Service
for support.
Additionally, the bottom row of buttons allows the configuration to be saved, read, loaded and stored.
Load Users from RMLD
will read the existing data from the RMLD CS and populate the System Users form
with that data. This is helpful to validate recent changes.
Download Users to File
will allow the System Users data to be saved to a PC as a JSON formatted text file.
It is highly recommended to Download Users to File for archival purposes.
Load Users From File
will allow a previously saved System Users file to populate the form and configure the
RMLD-CS. This is very useful when managing multiple instruments as only specific changes would then be
made from a template.
“
Save Users to RMLD
” must then be cli
cked to save the edits to the instrument.
Add User
allows for the creation of new users as shown in the next section.
Содержание HPN105354
Страница 1: ......