What if the instrument failed the Self-Test?
Most often, the causes are due to the instrument not being properly positioned in the case, the case was
moved during the test or the battery level is too low to perform the test. Ensure sufficient battery charge
before attempting the Self-Test procedure. Make sure the instrument is in its proper position, all the way
down, and flat. After rechecking the above, repeat the Self-Test procedure.
CAUTION: Should the instrument not pass after several attempts, do not use the instrument for survey
work until the problem is corrected. Contact HEATH for further assistance.
NOTE
: Laser wavelength drift is a normal characteristic of the RMLD-CS. The rate of drift is low and Self-
Test
will sweep the laser’s wavelength range to determine the optimum wavelength.
Survey work conducted
with an instrument which has drifted and was not Self-Tested regularly may need to be redone. If the
instrument is still not able to successfully perform a Self-Test after three attempts, contact HEATH for
assistance.
Audio Indications and Modes
Audio Warning Indications
A continuous alarm will indicate a warning and / or fault such as an infrared laser low signal level condition,
instrument fault or low battery capacity. The warnings and faults will also be displayed via pop-ups on the
display.
If low light, the operator needs to move in closer to get in range.
If over saturation, the operator should back up from the target or point the instrument at a less reflective
surface.
Should the warning(s) persist, it could be due to a fault in the instrument. Check the display and follow the
instructions in the troubleshooting guideline.