Settings
34
A02590_04_E00_00
HBM: public
TIM40
12.1.10 Transducer parameterization/Analog output
Select the signal at the current or voltage output (low‐pass 1 or 2) and specify
the scaling points with two values each (for voltages in the range -10 V to +10
V and for current 4 mA to 20 mA with reference to the selected torque measur
ing range).
1. Point
Torque
Voltage
-3,000 N
@
m
-10
+10
2. Point
Voltage range
"
10 V
Measuring range of torque transducer,
e.g. -3 kN
@
m to +3 kN
@
m
+3,000 N
@
m
12.1.11 Signal conditioning
You can connect a shunt and carry out a zero balance. We recommend zero
ing only in the range within
"
10% of the nominal (rated) measuring range.
Greater values are displayed in red in the browser.
12.1.12 Save/load parameters
You can save and load various settings in four parameter sets in the interface
module and reset all settings to the factory settings.
If you need more than four parameter sets, you can save these on the hard
drive of your PC (filename.tim40). Click on the link ”
Configuration file
” and con
firm the download into a directory of your choice.