Interface description
40
A02590_04_E00_00
HBM: public
TIM40
Meaning
Values
Bit
Meaning
Byte
6..7
00 = Parameter set 1
01 = Parameter set 2
10 = Parameter set 3
11 = Parameter set 4
Actual parameter
set
2
Error status
0..7
0
Reserved
RPDO1 control byte
CAN Identifier (COB ID)
512 (200h) + CAN address
Data
1. data byte
Control byte
Control word
Byte
Meaning
Bit
Values
Meaning
1
Control byte
0
0 = no function
1 = zeroing of actual
measured value
Torque zeroing
1
0
Reserved
2
0 = Shunt off
1 = Shunt on
Activate shunt
calibration
3..5
0
Reserved
6..7
00 = Parameter set 1
01 = Parameter set 2
10 = Parameter set 3
11 = Parameter set 4
Load parameter set.
Implemented when the
value has changed
13.1.5
Activation of PDO output
The exchange of cyclic PDOs only starts once the module has been brought to
the ”Operational” state. This is done with the ”Start_Remote_Node” message.
Switch all configured PDO's to ”operational”:
CAN Identifier
0
1. data byte
1 (01hex)
2. data byte
Module address (0 =, ID 110 = 6e hex)