CAUTION
Before starting a calibration or verification, it is recommended that the
ISOBE5600 system and the calibration standards run at least one hour to
achieve the best test specifications.
6.2.2
Starting the Verification test
When selecting bandwidth testing; there are now optional tests that can be
performed based on the depth of information required.
The basic Bandwidth test is performed by default and will calculate only the
-
3 db bandwidth value.
Figure 6.3:
Board Verification window (Bandwith tests)
A Full Bandwith
Selecting the
Full
check box:
This test is longer than the quick bandwidth test. It tests more input values
to give a better picture of the bandwidth curve before and after the
-
3db
value, A separate log file is produced.
B Mil standard
Selecting the
Mil
itary
standard
check box:
This will test all DC Gain input ranges possible and takes the most time.
When selecting
Mil
you will have the option to select the report to save and
display as a pre or post
Cal
ibration, these will then be separate files.