7.6
DC Output Test
General information
Required equipment:
l
Multimeter
Isolated Probe System settings:
l
The DC Output test can be done without the Transmitter connected.
Procedure:
Six voltages are generated by the ISOBE5600 Receiver one at a time. The
voltages are equally divided from
-
45 % to +45 % of Range and are denoted
as
Vout
i
.
These output voltages are measured with the Multimeter and denoted as
Vmeas
i
.
A “least squares regression” line is then calculated as follows:
DC Output Test
l
Gain Err
= (Slope
-
1) * 100 %
l
Offset Err
= (Intercept / Range) * 100 %
ISOBE5600
I2716-4.0 en
101