7.7
Output Noise Test
General information
Required equipment:
l
Multimeter
Isolated Probe System settings:
l
The Output Noise test can be done without the Transmitter connected.
Procedure:
The output is set to 0 V. The output AC voltage (rms) is measured with the
Multimeter:
Vmeas
rms
Noise
rms
=
Vmeas
rms
/ Range * 100 %
ISOBE5600
102
I2716-4.0 en