5 Calibration
5.1
Introduction
The calibration process contains the following tests depending on the selected
devices:
l
Transmitter
-
DC Calibration
l
Transmitter
-
AC Calibration
l
Receiver
-
DC Out Cal
CAUTION
Before starting a calibration or verification, it is recommended that the
ISOBE5600 system and the calibration standards run at least one hour to
achieve the best test specifications.
CAUTION
HBM uses state
-
of
-
the
-
art electronic components in its equipment. These
electronic components can be damaged by discharge of static electricity
(ESD). Therefore we must emphasize the importance of ESD preventions
when removing or installing boards.
Note
The calibration and verification tests depend on the installed and selected
board. Therefore the windows displayed may differ from the illustrations used
in this manual.
ISOBE5600
I2716-4.0 en
67