Calibration Process
The complete verification process is fully automated and delivers PASS / FAIL information with the press of a button.
The same applies to most of the calibration process, where electronic intelligence is used to retain the best performance possible. Most of this
is fully automated and no user interaction is needed to restore the modules to the best accuracy achievable.
Only in the rare case that AC bandwidth could be improved the use of manual user interaction is required. Then the manual and the software
itself help guide you through the needed steps. Exact process description and direct readouts make even manual procedures easy to work
with.
Figure A.3:
ISOBE5600 AC adjustment description and readouts
Verification Process
During the fully automated verification, no changes are made to any settings. The end result of the verification is a listing of all findings and an
overall PASS / FAIL result. The manual explains every verification process the software uses, including a detailed description of the calculation
methods to establish the specifications.
Supported verification steps
(Not every model requires all steps)
Voltage DC Gain, Offset, Linearity and MSE (Maximum Static Error)
AC Coupling
Input Bandwidth
Input Noise
CMRR (Common Mode Rejection Ratio)
DC Output Gain and Offset
Output Noise
ISOBE5600
106
I2716-4.0 en
Содержание Genesis ISOBE5600
Страница 1: ...Calibration and Verification Software Version 2 22 ISOBE5600 English User Manual I2716 4 0 en...
Страница 4: ...4 I2716 4 0 en...
Страница 40: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 ISOBE5600 40 I2716 4 0 en...
Страница 54: ...Figure 4 1 Connecting directly to your PC with the NI GPIB USB H USB controller ISOBE5600 54 I2716 4 0 en...
Страница 126: ...126 I2716 4 0 en...
Страница 127: ......