49
7 Specification
7.1 Input Specification
7.2 Output Specification
Input
(2 channels)
Pt100
(IEC751)
-200 ~ 640 ℃, ±0.1 % of FS
TC_K
-200 ~ 1370 ℃, ±0.1 % of FS
TC_J
-200 ~ 1200 ℃, ±0.1 % of FS
TC_E
-200 ~ 1000 ℃, ±0.1 % of FS
TC_T
-200 ~ 400 ℃, ±0.1 % of FS
TC_R
0 ~ 1700 ℃, ±0.15 % of FS
TC_S
0 ~ 1700 ℃, ±0.15 % of FS
mV
0 ~ 100 mV or -10 ~ 20 mV (-999.9 ~ 9999.9), ±0.1 % of FS
V DC
0 - 10 V (Signal Input Range setup is available, -999.9 ~ 9999.9),
±0.1 % of FS
4 - 20 mA Use 250 Ω external resistance, V DC Use after setup of V DC 1 - 5 V
Input Resolution
24 bit
Accuracy of indication
0.1 ℃
Sampling cycle
Each channel 500 msec
Input Resistance
More than 1 ㏁
Maximum allowable
resistance of line
Less than 50 Ω / Line (Resistance of lines is a same condition)
RJC
±1.5 ℃
Control
Output
(SCR)
4 - 20 mA DC(Resistive load : less than 600 Ω)
Output Resolution : 16 bit
Accuracy of Output : ±0.1 % of FS
Output Ripple : 0.2 % of FS
(SSR)
24 V DC Pulse (Resistive load : More than 600Ω)
Min. Pulse Range : 10 ms
Cycle Time : Available to select 1 - 1000 s
Relay Output
External Relay by using internal relay or TR output (Max.24 V, 100 mA)
Kind of Internal Relay : N/O -> 250 V AC 5 A / 30 V DC 5 A
N/C -> 250 V AC 2 A / 30 V DC 1 A
Transmission Output
(4 - 20 mA)
Avaliable to select PV/SV/H.MV/C.MV of each channel
Output Accuracy : ±0.1 % of FS
In case of heating/cooling control, if control output of heating is SCR or
SSR, this channel can not use transmission output.
Transmission Output
Period of outputrenewal
Each channel 500 ms