Table 2-62 Mask Setting function menu 1:
Option
Setting
Introduction
X Mask
xdiv
Turn the Universal knob to set the range
of horizontal clearance to the waveform.
<0.04div-4.00div>
Y Mask
ydiv
Turn the Universal knob to set the range
of vertical clearance to the waveform.
<0.04div-4.00div>
Create
Mask
Create a test mask according to the
above clearance.
Location
Internal
External
Select position to store created mask.
Next Page
Page 1/2
Press this button to enter the second
page of Mask Setting menu.
RUN Pass/Fail test, please follow next page:
●
Press
UTILITY button to enter the Utility menu.
●
Press
the Next Page Page1/4 option button.
●
Press
the Next Page Page2/4 option button to enter the third page
of Utility menu
●
Press
the Pass/Fail option button to enter the Pass/Fail function menu.
●
Press
the Enable Test option button to select On.
●
Press
the Source option button to select input signal channel.
●
Press
the Next Page Page1/2option button to enter the second page of
the Pass/Fail menu.
●
Press
the Mask Setting option button to enter the first page of Mask menu.
●
Press
the X Mask button; turn the Universal knob to adjust the horizontal
clearance.
●
Press
the Y Mask button; turn the Universal knob to adjust the vertical
clearance.
●
Press
the Create Mask button to create mask and you also can enter
the next page of Mask menu to recall the stored mask.
●
Enter
the second page of Pass/Fail function menu, press the Output
option button to set the output option.
●
Enter
the first page of Pass/Fail function menu, press the Operate
option button to select to run the pass/fail test.
Option
Setting
Introduction
Table 2-63 Mask Setting function menu 2:
Save
Store created mask setting.
Load
Recall stored mask setting.
Return
Return to the main menu of mask setting.
Last Page
Page 2/2
Return to the first page of the Mask Setting menu.