Picture 1-2 Display area
1. Trigger status
Armed:The oscilloscope is acquiring pre-trigger data. All triggers are ignored in
this state.
Ready: All pre-trigger data has been acquired and the oscilloscope is ready to
accept a trigger.
Trig'd: The oscilloscope has seen a trigger and is acquiring the posttrigger data.
Stop: The oscilloscope has stopped acquiring waveform data.
Auto: The oscilloscope is in auto mode and is acquiring waveforms in the
absence of triggers.
Scan: The oscilloscope is acquiring and displaying waveform data continuously
in scan mode.
2. Display the position of the present waveform window
3. Marker shows horizontal trigger position. Turn the HORIZONTAL POSITION
knob to adjust the position of the marker.
4. Print Key option is set to Print Picture.
Print Key option is set to Save Picture.
5. Back USB option is set to Computer.
Back USB option is set to Printer.
6. Show the waveform's channel position.
7. Signal Coupling symbol.
8. Readout shows the vertical factor of the channels.
9. "B" icon shows that bandwidth function is open.
10. Readout shows the main time base setting.
11. Icon shows the trigger type for the triggering
12. Readout shows the horizontal position.
13. Readout shows the trigger voltage.
14. Readout shows present signal frequency.
When you check whether or not the oscilloscope could work smoothly,
please operate as following:
1.2 Function checking
1. Power On the oscilloscope.Press DEFAULT
SETUP to show the result of the self check.
The probe default attenuation is 1X.
2. Set the switch to 1X on the probe and connect
the probe to channel 1 on the oscilloscope. To
do this, align the slot in the probe connector
with the key on the CH 1 BNC, push to connect,
and twist to the right to lock the probe in place.
Connect the probe tip and reference lead to the
PROBE COMP connectors
3. Press AUTO to show the 1 KHz frequency
and about 3V peak-peak square wave in
couple seconds
4. Press CH1 two times to cancel the channel 1,
PressCH2 to change screen into channel 2,
reset the channel 2 as step 2 and 3.
Probe
Comp
02
Digital
S
torage
O
scilloscope
Digital Storage Oscilloscope
03
Picture 1-3
1.Pass/Fail Output
2.RS-232 Connector
3.Back USB Connector
4.Power Input Connector