Note 

1.  The oscilloscope vertical response rolls off slowly above its bandwidth , which 
     depending on the model, or 20 MHz when the Bandwidth Limit option isset to On. 
     Therefore, the FFT spectrum can show valid frequency information higher than the 
     oscilloscope bandwidth. However, the magnitude information near or above the 
     bandwidth will not be accurate.
2.  If the channel is set to DC coupling,you can quickly measure the DC component of 
     the signal by simply nothing its distance from the ground symbol
3.  If the channel is set to AC coupling,the DC component of the signal is blocked allowing 
     you to use greater sensitivity to display the AC component of the signal..

     

     GND

 Coupling: Use GND coupling to display a zero-volt waveform. 

            Internally, the channel input is connected to a zero-volt reference level.
     

    

 Fine Resolution:The vertical scale readout displays the actual Volts/Div 

            setting while in the fine resolution setting. Changing the setting to coarse 
            does not change the vertical scale until the VOLTS/DIV control is adjusted.
     

     Wave

 Cancel:Press the menu button for the channel to display its vertical 

            menu. Push the menu button again to remove the waveform.

Setting up the CH1、CH2 Channels 

    1.   Set up the channel couple

            Take the CH1 for example; the tested signal is a sine wave signal with DC 
            deflection: 
     

     PressCH

1→Coupling→AC, Set to AC couple mode. It blocks the DC 

            component of the input signal.      
     

     PressCH

1→Coupling→DC, Set to DC couple mode. Both

            DC and AC component could be obstructed. 
     

     PressCH

1→Coupling→GND, Set to GROUND mode. It disconnects 

            the input signal. 

● Set to AC coupling

● AC Status

○ Set to DC coupling

○ DC Status

○ Set to GND coupling

○ GND Status

Table 2-5 Digital Filter function menu:

Digital Filter

On
Off

Turn on the digital filter.
Turn off the digital filter.

Type

Setup as LPF(Low Pass Filter).
Setup as HPF(High Pass Filter).

Setup as BPF(Band Pass Filter).
Setup as BRF(Band Reject Filter).

Upper_limit

Turn the Universal knob to set upper limit. 

Lower_limit

Turn the Universal knob to set lower limit.

Return

Return the digital filter main menu.

Option

Setting

Instruction

Table 2-4 CH1 & CH2 function menu 2:

Invert

On

Off

Turn on invert function.

Turn off invert function.

Input

1MΩ

Setup 1MΩ as input impedance 

Digital Filter

Press this button to enter the Digital Filter 
menu.(See table 2-5)

Next Page

Page 2/2

Press this button to return the first page menu.

Option

Setting

Instruction

Each channel has its own separate Menu. The items are set up separately 
according to each channel.

    2.   Set up the channel bandwidth limit

            Take the CH1 for example; the tested signal is a pulse signal with the high
            frequency surge:

     

     Press

 Ch1→BW Limit→ On,Set the band width Limited to open state. 

            The high frequency component which is higher than 20MHz obstructed. 
     

     PressCH

1→BW Limit→ Off, Set bandwidth Limited to close state, 

            the High Frequency component in the tested signal could pass. 

12

        

Digital

 S

torage

 O

scilloscope

  

  

Digital Storage Oscilloscope

        

13

        

  

  

Содержание 500MSa/s

Страница 1: ...lash drive save recall function Cursor measure covers Manual mode Track mode and Auto mode Channel waveform and its FFT waveform display on split screen Waveform Intensity and Grid Brightness can be a...

Страница 2: ...duct manual for further ratings information before making connections to the product Do Not Operate Without Covers Do not operate this product with covers or panels removed Use Proper Fuse Use only th...

Страница 3: ...l contains operating information for the Digital Storage Oscilloscopes The manual consists of the following chapters Digital Storage Oscilloscope Digital Storage Oscilloscope Catalogue Digital Storage...

Страница 4: ...ion Matching probes attenuation coefficient Digital Storage Oscilloscope Digital Storage Oscilloscope 01 2 12 Storage System 70 2 13 Utility System 83 2 13 1 System status 87 2 13 2 Language 87 2 13 3...

Страница 5: ...6 Show the waveform s channel position 7 Signal Coupling symbol 8 Readout shows the vertical factor of the channels 9 B icon shows that bandwidth function is open 10 Readout shows the main time base...

Страница 6: ...firmly inserting the tip onto the probe 2 Attach the probe tip to the PROBE COMP 3V connector and the reference lead to the PROBE COMP Ground connector Display the channel and then push the AUTO butt...

Страница 7: ...ion is set to Off after leaving the Cursor Menu but are not adjustable DISPLAY Display the Display Menu UTILITY Display the Utility Menu DEFAULT SETUP Recall the factory setup HELP Enter the online he...

Страница 8: ...o button See picture 2 3 Table 2 2 Auto set the function item Acquire Mode Adjusted to Sampling Display Format Y T Display Type Set to Dots for a video signal set to Vectors for an FFT spectrum otherw...

Страница 9: ...to save or recall them and you can adjust the Universal knob to select menu option 2 6 Vertical System Showing as follow picture the buttons and knobs are useful in the vertical system Volt div knob...

Страница 10: ...component of the input signal PressCH1 Coupling DC Set to DC couple mode Both DC and AC component could be obstructed PressCH1 Coupling GND Set to GROUND mode It disconnects the input signal Set to A...

Страница 11: ...s It will be helpful when you need to adjust the waveform vertical size in smooth steps Set to Coarse Set to Fine 4 Probe attenuation set In order to assort the attenuation coefficient you need to res...

Страница 12: ...hows the results after and FFT operation of the CH1 and CH2 Press the MATH button to display the waveform math operations Press the MATH button again to remove the math waveform display Table 2 6 MATH...

Страница 13: ...xt Page Page 1 2 Enter the second page of FFT menu To use the Math FFT mode you need to perform the following tasks 1 Set up the source time domain waveform Press the AUTO button to display a YT wavef...

Страница 14: ...FT window instruction Function Setting Introduction Rectangular Best frequency resolution worst magnitude resolution This is essentially the same as no window Symmetric transients or bursts Equal ampl...

Страница 15: ...ess CH1 button to display CH1 menu Turn the S div knob to adjust sampling rate at least double bigger than frequency of input signal If FFT display on full screen press CH1 button again to remove chan...

Страница 16: ...ce waveform to the chosen reference location REFA REFB On Off Recall the reference waveform on the screen Turn off the reference waveform Press the Ref button to display the Reference waveform menu Op...

Страница 17: ...ses the waveform to expand or contract around the screen center Horizontal POSITION knob 1 Adjust the horizontal position of all channels and math waveforms the position of the trigger relative to the...

Страница 18: ...If waveform acquisition is stopped using the RUN STOP or SINGLE button turn the S div knob to expand or compress the waveform 2 Select the horizontal time div scale factor for the main or the window...

Страница 19: ...he events that led up the trigger point everything to the right of the trigger point is called posttrigger information the amount of delay range pre trigger and posttrigger information available is de...

Страница 20: ...want the oscilloscope to acquire a single waveform press the SINGLE button Set up Enter the Trigger Setup Menu See table 2 14 32 Digital Storage Oscilloscope Digital Storage Oscilloscope 33 1 Set up T...

Страница 21: ...he Set Pulse Width option Set Width 20 0ns 10 0s Selecting this option can turn the universal to set up the pulse width Next Page Page 1 2 Press this button to enter the second page 34 Digital Storage...

Страница 22: ...nu 36 Digital Storage Oscilloscope Digital Storage Oscilloscope 37 Table 2 19 Functional Manu of Video Trigger 2 Option Setting Introduction Type Video When you select the video type put the couple se...

Страница 23: ...inted line number 4 Set up Standard Press the Next Page Page 2 2 option button Press the Standard option button to select PAL SECAM or NTSC 38 Digital Storage Oscilloscope Digital Storage Oscilloscope...

Страница 24: ...serve two irrelative signals at the same time You can select different trigger types for two vertical signals and selected types cover edge pulse video and slope trigger Trigger information of two cha...

Страница 25: ...ess this button to return the first page Option Setting Introduction Type Alternative The trigger signal comes from two vertical channels when you use alternative trigger In this mode you can observe...

Страница 26: ...el 1 and channel 2 Press the AUTO button Press the TRIG MENU button to enter trigger menu Press the Type option button to select Alternative Press the Source option button to select Ch1 Press the S di...

Страница 27: ...er during the holdoff time For a pulse train you can adjust the holdoff time so the oscilloscope triggers only on the first pulse in the train If you want to change holdoff time please follow next ope...

Страница 28: ...nes how the signal is digitized and the time base setting affects the time span and level of detail in the acquisition Advantage You can use this mode to reduce random noise Disadvantage This mode doe...

Страница 29: ...while waiting for the trigger condition to occur Detect the trigger condition Continue to acquire data until the waveform record is full Display the newly acquired waveform Base time The oscilloscope...

Страница 30: ...em The display function could be expressed by the DISPLAY Button Table 2 30 Display system function menu 1 Option Setting Introduction Type Vectors Dots Vectors fills the space between adjacent sample...

Страница 31: ...lay mode Grid Display grids and axes on the screen Turn off the grids Turn off the grids and axes Menu Display 2sec 5sec 10sec 20sec Infinite Set menu display time on screen Next Page Page 2 3 Press t...

Страница 32: ...lloscopes restrict sampling rate at 1MSa s support 5KSa s 200MSa s 1 2 5 5 sequence adjusted The other series support 25KSa s 500MSa s 1 2 5 5 sequence adjusted 4Kpts memory depth Series Operate Intro...

Страница 33: ...plays the voltage in relation to time and test the wave form displayed There are scale Cursor and auto measure modes 2 11 1 Scale Measurement This method allows you to make a quick visual estimate For...

Страница 34: ...iversal knob to move cursor1 and cursor2 They could be moved when the corresponding cursor option are selected and cursor value will display on the bottom left and top left of the screen when you move...

Страница 35: ...al knob to move Cursor B horizontally The measurement values are displayed on the left of the top corner A T The horizontal position of Cursor A Time cursor centered around the midpoint of screen A V...

Страница 36: ...Cycle Vrms Vrms ROVShoot FOVShoot RPREShoot FPREShoot Display the corresponding icon and measure value of your selected Voltage measure parameter Return Return to the first page of auto measurement me...

Страница 37: ...ig and Vlow of a waveform The arithmetic mean over the first cycle in the waveform The arithmetic mean over the entire waveform The true Root Mean Square voltage over the first cycle in the waveform T...

Страница 38: ...rce option button to select CH1 or CH2 according to input signal channel Press the Type option button to select the parameter type that you want to measure The corresponding icon and value will displa...

Страница 39: ...ctures Select operating object Create new needed folder Delete present folder Select operating object Create new files in present directorys Delete present file Recall stored files or present settings...

Страница 40: ...e Save to option button to select Device Press the Setup option button to select No 1 Press the DISPLAY button to enter the Display menu Press the Type option button to select Dots Press the SAVE RECA...

Страница 41: ...Storage Recall waveforms in the oscilloscope Save to Device Save waveforms to the oscilloscope s internal memorizer waveform No 1 to No 20 Press the waveform option button or turn the Universal knob...

Страница 42: ...the oscilloscope and wait that the oscilloscope has initialized USB flash drive about ten seconds Press the Save to option button to select File Press the Save option button then you ll go into the Sa...

Страница 43: ...then you ll go into the Save Recall interface Create a file then press the Confirmbutton about ten seconds there is a prompting message that Save data success displaying on the screen now waveform da...

Страница 44: ...scope can display reference waveforms at a time Table 2 49 Picture storage function menu Option Setting Introduction Type Pictures Menu for the Storage Recall waveform interface image Print Key Print...

Страница 45: ...ve seconds there is a prompting message that Save data success displaying on the screen now CSV file has been saved to USB flash drive 2 13 Utility System UTILITY The Utility System Button Table 2 51...

Страница 46: ...hen you execute EasyScope software please select Computer Now the computer icon displays on the screen Next Page Page 2 4 Press this button to enter third page Table 2 53 Functional Menu of the Utilit...

Страница 47: ...display a series of information about this oscilloscope Table 2 56 System status instruction Option Introduction Startup Times List the boot strap times Software version List the software version Hard...

Страница 48: ...rmal Draft Fine ID Print Default On Off Next Page Page 2 2 Press this button to enter first page of print setup menu Displays Settings available on your PictBridge compatible printer Table 2 57 Print...

Страница 49: ...at the most measurement precision You can run this procedure at any time If the operating temperature changes by more than 5 C or the units run less than thirty minutes you should do the self calibrat...

Страница 50: ...s the Single for test 2 Keyboard Test Select keyboard Test to enter the key test interface the on screen lathy rectangle shapes represent the front panel keys The shapes with two arrows beside them re...

Страница 51: ...ging the input signal is within the predefined mask or not Table 2 61 Pass Fail function menu 2 Option Setting Introduction Output Pass Fail Output when pass condition detected Output when fail condit...

Страница 52: ...the Source option button to select input signal channel Press the Next Page Page1 2option button to enter the second page of the Pass Fail menu Press the Mask Setting option button to enter the first...

Страница 53: ...e select the Universal knob to adjust the max record frame Press in Operateoption to record waveform Play Back Play back current record waveforms or saved record waveforms Table 2 65 Waveform play bac...

Страница 54: ...Return Press to return the waveform recorder main menu Next Page Page 2 2 Return to the first page of Play Back function menu 100 Digital Storage Oscilloscope Digital Storage Oscilloscope 101 2 13 9...

Страница 55: ...n buttons to see its help information Table 2 69 Recorder s Setting Menu Option Setting Description Viewer Full Screen Split Waveform of Full screen recording and channel recalling Waveform of recordi...

Страница 56: ...ton to enter the Time Measure menu Press the Source option button to select input signal channel Press the Type option button to select Freq The corresponding icon and measure value will display on th...

Страница 57: ...Menu Under this menu set the trigger type to edge Edge type is Rising source is CH1 trigger mode is Single coupling is DC Adjust the horizontal time base and vertical scale to the satisfied range Turn...

Страница 58: ...urn the Volts Div knob to ensure that the entire video signal is visible on the screen 3 5 2 Triggering on Video Lines 3 6 Application of X Y function Viewing Impedance Changes in a Network Connect th...

Страница 59: ...ontal position is at a limit when you turn the horizon position knob Volts Div at limit Mention you that the vertical voltage have already touched the Min 2MV div or the Max 5V div Volts position at l...

Страница 60: ...ger mode Attempt to change the coupling into HF Reject or LF Reject display so that the High low frequency noise disturb the trigger should be filtrated 5 Press RUN STOP button but no display Check th...

Страница 61: ...cifications the oscilloscope must first meet the following conditions The oscilloscope must have been operating continuously for thirty minutes within the specified operating temperature You must perf...

Страница 62: ...MHZ 40 when using probe X1 Measure System Auto Measure 32 Types Vpp Vmax Vmin Vamp Vtop Vbase Vavg Mean Crms Vrms ROVShoot FOVShoot RPREShoot FPREShoot Rise time Fall time Freq Period Wid Wid Dut Dut...

Страница 63: ...Operating 3000m Not operating 15 266m X Y Mode X pole Input Y Pole Input Channel 1 CH1 Channel 2 CH2 Phase Error 3 degrees Sample Frequency XY mode has a breakthrough that trad oscilloscopes restrict...

Страница 64: ...oom Scale Display CH1 Hanning 1X dBVrms Split Standard Accessories CURSOR Type Off Source CH1 Horizontal voltage 3 2divs Vertical time 5divs MEASURE Source CH1 Type average ACQUIRE three mode options...

Страница 65: ...re the LCD display will be exposed to direct sunlight for long periods of time CAUTION To avoid damage to the instrument or probes do not expose them to sprays liquids or solvents Cleaning If this ins...

Отзывы: