11
ALK
ALINITY OF W
ATER
ALKALINITY OF WATER
0 to 2500 mg/L CaCO
3
, pH 4.5 Endpoint
METHOD ID:
HI1004EN
DESCRIPTION
Method for the determination of total (methyl red) alkalinity in
water by titration of a sample to pH 4.5. The results are expressed in
mg/L (ppm) as calcium carbonate
.
For the determination of phenolphthalein alkalinity, set the endpoint
to pH 8.3.
REFERENCE
Standard Methods for the Examination of Water and Wastewater
21
st
edition, Method 2320B
ELECTRODE
• HI1131B Combination pH Electrode
• HI7662-T Temperature Probe
REAGENTS
• HI70463 0.1N Hydrochloric Acid (1 L)
• HI70436 Deionized Water (1 gal)
ACCESSORIES
• HI70300L Storage Solution (500 mL)
• HI7082
Electrode Fill Solution (4 x 30 mL)
• HI7004L pH 4.01 Buffer Solution (500 mL)
• HI7007L pH 7.01 Buffer Solution (500 mL)
• HI7010L pH 10.01 Buffer Solution (500 mL)
• HI740036P 100 mL Plastic Beaker (10 pcs)
• 50 mL Class A Volumetric Pipette
DEVICE PREPARATION
• Connect the pH electrode and temperature probe to the titrator.
• Install a 25 mL burette filled with 0.1N hydrochloric acid
(HI70463) on pump one and verify that no air bubbles are
present in the burette or tubing. If necessary prime the burette
until all the air has been removed completely.
• For the determination of the exact concentration of the 0.1N
hydrochloric acid, follow
HI0002EN 0.1N Hydrochloric Acid
Titrant Concentration.
• Press
from the main screen. Use the arrow keys to
highlight
Alkalinity of Water
and press
.
ELECTRODE PREPARATION
• Press
from the main screen, if necessary select the
analog board and press
.
• Calibrate the electrode using pH 4.01, 7.01 and 10.01 buffers.
Refer to the instruction manual for calibration procedure.
SAMPLE PREPARATION
• Use a Class A volumetric pipette to transfer exactly 50.00 mL of
sample to a clean 100 mL plastic beaker.
ANALYSIS
• Place the beaker under the stirrer assembly and lower it to
immerse the pH electrode, temperature sensor and stirrer.
Ensure that the reference junction of the pH electrode is 5 to 6
mm below the surface. If necessary add extra deionized water.
Note
:
The dispensing tip should be slightly submerged in the
sample.
• Press
, the titrator will start the analysis.
• At the end of the titration, when pH 4.50 is reached, “Titration
Completed” will appear with the result. The result is expressed
in
mg/L as calcium carbonate
.
• Remove the pH electrode, temperature probe and stirrer from the
sample and rinse them thoroughly with deionized water.
• Record the result.
Содержание HI931
Страница 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Страница 2: ......
Страница 4: ......
Страница 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Страница 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Страница 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Страница 20: ......
Страница 24: ...1 2 ...
Страница 26: ...2 2 ...
Страница 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Страница 36: ...2 12 ...
Страница 38: ...3 2 ...
Страница 46: ...4 2 ...
Страница 92: ...5 34 ...
Страница 94: ...6 2 ...
Страница 98: ...6 6 ...
Страница 100: ...7 2 ...
Страница 112: ...7 14 ...
Страница 114: ...8 2 ...
Страница 120: ...8 8 ...
Страница 122: ...9 2 ...
Страница 134: ...9 14 ...
Страница 136: ...10 2 ...
Страница 144: ...10 10 ...
Страница 146: ...11 2 ...
Страница 153: ...A1 1 TECHNICAL SPECIFICATIONS APPENDIX 1 TECHNICAL SPECIFICATIONS A1 HI931 TECHNICAL SPECIFICATIONS A1 3 ...
Страница 154: ...A1 2 ...
Страница 158: ...A2 2 ...
Страница 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Страница 170: ......
Страница 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Страница 196: ...2 ...
Страница 198: ...4 ...
Страница 216: ...22 THEORY931 932 10 18 ...
Страница 217: ......