9
QUICK START GUIDE
HOW TO CALIBRATE A pH ELECTRODE
To enter pH calibration mode, press
then
to enter pH mode. Once in pH
mode, press
to enter pH calibration mode.
PREPARATION
Pour small quantities pH 4.01, pH 7.01 and pH 10.01 buffer solutions into clean beakers. If
possible, use plastic beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two beakers for each
buffer solution: one for rinsing the electrode and one for calibration.
CALIBRATION PROCEDURE
Three buffer entry types are available: Automatic, Semi-automatic and Manual Selection.
The default option is Manual Selection.
• If the instrument has been previously calibrated and calibration was not cleared, the old
calibration can be cleared by pressing
.
Note:
It is very important to clear calibration history when a new electrode is used
because most errors and warning messages that appear during calibration depend
on calibration history.
• Use the
or
to select pH 4.01 buffer solution.
• Use the second beaker of pH 4.01 buffer solution to rinse the pH electrode, temperature
probe and propeller stirrer.
• Immerse the pH electrode, temperature probe and propeller stirrer in the pH 4.01 buffer
solution. The pH electrode’s bulb must be completely immersed in the buffer solution and
the reference junction needs to be 5-6 mm below the surface. Add additional buffer if
necessary.
• Press
to turn on the propeller stirrer.
• Once the reading has stabilized, press
to update the calibration.
• Repeat this procedure for pH 7.01 and 10.01 buffer solutions.
• Press
to accept and exit pH calibration mode.
Содержание HI 902 Color
Страница 6: ...6 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View sale ponpe com www ponpe com ...
Страница 18: ...18 QUICK START GUIDE QS 902C 10 10 sale ponpe com www ponpe com ...
Страница 19: ...1 INSTRUCTION MANUAL HI 902 Color AUTOMATIC POTENTIOMETRIC TITRATOR Revision 2 00 www ponpe com sale ponpe com ...
Страница 22: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Страница 24: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Страница 26: ...2 2 SETUP sale ponpe com www ponpe com ...
Страница 29: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Front View 2 3 2 Titrator Rear View www ponpe com sale ponpe com ...
Страница 36: ...3 2 USER INTERFACE sale ponpe com www ponpe com ...
Страница 44: ...4 2 GENERAL OPTIONS sale ponpe com www ponpe com ...
Страница 100: ...5 44 METHODS sale ponpe com www ponpe com ...
Страница 102: ...6 2 TITRATION MODE sale ponpe com www ponpe com ...
Страница 106: ...6 6 TITRATION MODE sale ponpe com www ponpe com ...
Страница 138: ...8 2 AUXILIARY FUNCTIONS sale ponpe com www ponpe com ...
Страница 148: ...9 2 MAINTENANCE PERIPHERALS sale ponpe com www ponpe com ...
Страница 159: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3 sale ponpe com www ponpe com ...
Страница 160: ...A1 2 APPENDIX 1 sale ponpe com www ponpe com ...
Страница 164: ...A2 2 APPENDIX 2 sale ponpe com www ponpe com ...
Страница 170: ...A3 2 APPENDIX 3 sale ponpe com www ponpe com ...
Страница 174: ...A3 6 APPENDIX 3 sale ponpe com www ponpe com ...
Страница 176: ...A4 2 APPENDIX 4 sale ponpe com www ponpe com ...
Страница 184: ...A5 2 APPENDIX 5 sale ponpe com www ponpe com ...
Страница 193: ...A5 11 APPENDIX 5 ...
Страница 194: ...A5 12 APPENDIX 5 MAN902C 10 10 ...
Страница 196: ...1 2 INTRODUCTION ...
Страница 223: ...1 TITRATION THEORY TITRATION THEORY HI 901 and HI 902 AUTOMATIC POTENTIOMETRIC TITRATOR sale ponpe com www ponpe com ...
Страница 224: ...2 TITRATION THEORY sale ponpe com www ponpe com ...
Страница 248: ...26 TITRATION THEORY Titr THEORY 901 902 04 10 ...