Method ID:
HI1011EN
Troubleshooting 1
Titrator Application Methods:
General Methods - r. 2.3
Temperature (°C)
FACTOR
17.0
18.0
19.0
20.0
21.0
22.0
23.0
24.0
25.0
26.0
27.0
28.0
29.0
30.0
1.002290
1.002467
1.002654
1.002853
1.003061
1.003282
1.003512
1.003752
1.004002
1.004261
1.004531
1.004809
1.005097
1.005395
The specifications of the dosing accuracy is
0.1%
of full burette volume (
0.1% * 25 mL =
0.025
mL). For the accuracy of other burette volumes, see
the HI 901 / HI 902 Instruction Manual.
If the results are not correct, check all fittings for
leakage, and burette and tubing for air bubbles.
Repeat the measurements.
The table below contains the settings to be used for
other burette volumes:
Burette
Volume
Dosing
Type
Pre-Titr.
Volume
Max. Titr.
Volume
5 mL
10 mL
25 mL
4.000 mL
8.000 mL
20.000 mL
4.000 mL
8.000 mL
20.000 mL
8.000 mL
16.000 mL
40.000 mL
Temperature Channel Fast Check Procedure:
Connect the shorting cap to the BNC socket on
Analog Board 1.
Connect the HI 762000C 0°C temperature key to
the RCA socket (temperature sensor input) on
Analog Board 1.
Once on the main screen select “
Mode
”, ensure
that Analog Board 1 is active then select “
mV 1
”.
Press “
General Options
”, use the arrow keys to
highlight ‟Temperature‟ and press “
Select
”.
Select „Temperature Source‟ and then „Automatic
Temperature‟.
Press “
Escape
” two times to return to the main
screen.
The titrator should display ATC 0.0
0.4°C with
no fluctuation or drift.
Connect the HI 762070C 70°C temperature key to
the RCA socket (temperature sensor input) on
Analog Board 1
The titrator should display ATC 70.0
0.4°C with
no fluctuation or drift.
Note:
This procedure can be repeated on analog 2
Temperature & mV Channel Logging Procedure
:
Connect the shorting cap to the BNC socket on
Analog Board 1.
Connect the HI 762000C 0°C temperature key to
the RCA socket (temperature sensor input) on
Analog Board 1.
Once on the main screen select “
Mode
”, ensure
that Analog Board 1 is active then select “
mV 1
”.
Pr
ess “
General Options
”, use the arrow keys to
highlight ‟Temperature‟ and press “
Select
”.
Select „Temperature Source‟ then „Automatic
Temperature‟.
Press “
Escape
” two times to return to the main
screen.
Press “
mV Setup
”, use the arrow keys to highlight
‟Logging Interval‟. Set the logging interval to 15
seconds and press “
Accept
”. Press “
Escape
” to
return to the main screen.
Press the “result” key and use the arrow keys to
highlight ‟Setup pH/mV/ISE Report‟. Press
“
Select
”.
Select „Potential‟ and 'Temperature and Units‟ (the
selected fields are marked with an “
*
”). All other
fields should be unselected.
Press “
Save Report
” to return to the Data
Parameters screen.
Press “
Escape
” to return to the main screen. Once
on the main screen, select “
Mode
” then “
mV 1
” to
enter mV mode on Analog Board 1.
Press “
Start Log
” to start the automatic log.
Let the log run for about 10 minutes. Press “
Stop
Log
” to stop the automatic log.
Press “
Result
”, highlight „Review Available
Reports‟, and press “
View Report”.
The mV column should display 0.0
0.1 mV, and
the temperature column should display 0.0°C
0.4°C.
This procedure can be repeated using the HI
762070C 70°C temperature key.
Note:
This procedure can be repeated on analog 2
If any problems occur, please contact your nearest
Hanna Service Office.
Содержание HI 902 Color
Страница 6: ...6 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View sale ponpe com www ponpe com ...
Страница 18: ...18 QUICK START GUIDE QS 902C 10 10 sale ponpe com www ponpe com ...
Страница 19: ...1 INSTRUCTION MANUAL HI 902 Color AUTOMATIC POTENTIOMETRIC TITRATOR Revision 2 00 www ponpe com sale ponpe com ...
Страница 22: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Страница 24: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Страница 26: ...2 2 SETUP sale ponpe com www ponpe com ...
Страница 29: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Front View 2 3 2 Titrator Rear View www ponpe com sale ponpe com ...
Страница 36: ...3 2 USER INTERFACE sale ponpe com www ponpe com ...
Страница 44: ...4 2 GENERAL OPTIONS sale ponpe com www ponpe com ...
Страница 100: ...5 44 METHODS sale ponpe com www ponpe com ...
Страница 102: ...6 2 TITRATION MODE sale ponpe com www ponpe com ...
Страница 106: ...6 6 TITRATION MODE sale ponpe com www ponpe com ...
Страница 138: ...8 2 AUXILIARY FUNCTIONS sale ponpe com www ponpe com ...
Страница 148: ...9 2 MAINTENANCE PERIPHERALS sale ponpe com www ponpe com ...
Страница 159: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3 sale ponpe com www ponpe com ...
Страница 160: ...A1 2 APPENDIX 1 sale ponpe com www ponpe com ...
Страница 164: ...A2 2 APPENDIX 2 sale ponpe com www ponpe com ...
Страница 170: ...A3 2 APPENDIX 3 sale ponpe com www ponpe com ...
Страница 174: ...A3 6 APPENDIX 3 sale ponpe com www ponpe com ...
Страница 176: ...A4 2 APPENDIX 4 sale ponpe com www ponpe com ...
Страница 184: ...A5 2 APPENDIX 5 sale ponpe com www ponpe com ...
Страница 193: ...A5 11 APPENDIX 5 ...
Страница 194: ...A5 12 APPENDIX 5 MAN902C 10 10 ...
Страница 196: ...1 2 INTRODUCTION ...
Страница 223: ...1 TITRATION THEORY TITRATION THEORY HI 901 and HI 902 AUTOMATIC POTENTIOMETRIC TITRATOR sale ponpe com www ponpe com ...
Страница 224: ...2 TITRATION THEORY sale ponpe com www ponpe com ...
Страница 248: ...26 TITRATION THEORY Titr THEORY 901 902 04 10 ...