7-14
pH, mV & ISE MODE
In ATC mode, the pH value can be modified by pressing
.
If the Automatic buffer entry type was selected for the calibration procedure, the
instrument will automatically select the closest buffer to the measured pH value from
the edit buffer group (see section 7.1.2.4 Buffer Group Edit).
If the Semiautomatic buffer entry type was selected for the calibration procedure, the
instrument will automatically select the closest buffers to the measured pH value from
all the available buffers and the buffer value can be selected with
or
.
CALIBRATION MESSAGES:
•
Wrong Buffer. Please check the buffer
: This message appears when the difference
between the pH reading and the value of the selected calibration buffer is significant. If
the message is displayed, check if you have selected the appropriate calibration buffer.
•
Wrong buffer temperature
: This message appears if the buffer temperature is out
of the defined temperature range.
•
Clean the electrode or check the buffer. Press
to update calibration
:
This message alerts the user that some dirt or deposits could be on the electrode.
•
Slope too low. Please check the buffer
: This message appears if the current
slope is under 80% or over 110% of default slope. Recalibrate the instrument using
fresh buffers.
•
Slope too high. Press
to clear the old calibration
: This message appears
as a result of an erroneous slope condition. Follow displayed instructions.
Содержание HI 902 Color
Страница 6: ...6 QUICK START GUIDE TITRATOR CONNECTIONS Front View Rear View sale ponpe com www ponpe com ...
Страница 18: ...18 QUICK START GUIDE QS 902C 10 10 sale ponpe com www ponpe com ...
Страница 19: ...1 INSTRUCTION MANUAL HI 902 Color AUTOMATIC POTENTIOMETRIC TITRATOR Revision 2 00 www ponpe com sale ponpe com ...
Страница 22: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Страница 24: ...1 2 INTRODUCTION sale ponpe com www ponpe com ...
Страница 26: ...2 2 SETUP sale ponpe com www ponpe com ...
Страница 29: ...2 5 SETUP 2 3 Installation 2 3 1 Titrator Front View 2 3 2 Titrator Rear View www ponpe com sale ponpe com ...
Страница 36: ...3 2 USER INTERFACE sale ponpe com www ponpe com ...
Страница 44: ...4 2 GENERAL OPTIONS sale ponpe com www ponpe com ...
Страница 100: ...5 44 METHODS sale ponpe com www ponpe com ...
Страница 102: ...6 2 TITRATION MODE sale ponpe com www ponpe com ...
Страница 106: ...6 6 TITRATION MODE sale ponpe com www ponpe com ...
Страница 138: ...8 2 AUXILIARY FUNCTIONS sale ponpe com www ponpe com ...
Страница 148: ...9 2 MAINTENANCE PERIPHERALS sale ponpe com www ponpe com ...
Страница 159: ...A1 1 APPENDIX 1 Appendix 1 Contents A1 TECHNICAL SPECIFICATIONS A1 3 sale ponpe com www ponpe com ...
Страница 160: ...A1 2 APPENDIX 1 sale ponpe com www ponpe com ...
Страница 164: ...A2 2 APPENDIX 2 sale ponpe com www ponpe com ...
Страница 170: ...A3 2 APPENDIX 3 sale ponpe com www ponpe com ...
Страница 174: ...A3 6 APPENDIX 3 sale ponpe com www ponpe com ...
Страница 176: ...A4 2 APPENDIX 4 sale ponpe com www ponpe com ...
Страница 184: ...A5 2 APPENDIX 5 sale ponpe com www ponpe com ...
Страница 193: ...A5 11 APPENDIX 5 ...
Страница 194: ...A5 12 APPENDIX 5 MAN902C 10 10 ...
Страница 196: ...1 2 INTRODUCTION ...
Страница 223: ...1 TITRATION THEORY TITRATION THEORY HI 901 and HI 902 AUTOMATIC POTENTIOMETRIC TITRATOR sale ponpe com www ponpe com ...
Страница 224: ...2 TITRATION THEORY sale ponpe com www ponpe com ...
Страница 248: ...26 TITRATION THEORY Titr THEORY 901 902 04 10 ...