MilkyWay
MilkyWay Handbook • Version 3 - 20140616 © Handicare AB
4: Rail systems with full room coverage - (x-y)/traverse systems / 31
17) Make additional test runs with the cross-rail. Observe the trolleys closely from the end. If the runs are not smooth
and without friction, alignment of the fixed rails must be checked again.
18) Move the endstop of the cross-rail to the desired positions, not necessarily to the end of the rail. Limitations in
the movement can be an advantage e.g. to avoid colliding with cupboards or the like.
NB! It is important that the endstops of the trolleys are mounted exactly opposite to enable the trolleys to stop
simultaneously.
19) Tighten all nuts and bolts.
20) Perform a final test of all functions including weight testing, see 7: Weight testing procedure.
21) Instruct the user and/or attendant on how to use the system.