background image

Änderungen vorbehalten / Subject to change without notice

18

Addressed messages

These commands are received by only a single device, which is addressed.

GTL

Switches to local

Go To Local

(manual) operation

SDC

Initializes an instrument

Selected Device Clear
PPC

Transmits the parallel poll byte (followed by PPE)

Parallel PolI Configure
GET

Activates a device function

Group Execute Trigger
TCT

Passes control to another controller

Take Control

Universal (unaddressed) messages

These commands are received by all devices on the bus that have Listener capability.

LLO

Prevents manual operation

Local Lockout
DCL

Initializes instruments

Device Clear
PPU

Removes parallel pollstatus

Parallel Poll Unconfig.
SPE

Initiates serial poll

Serial Poll Enable
SPD

Terminates serial poll

Serial Poll Disable

Secondary commands

These are used to transfer the parallel poll byte following PPC, for clearing the parallel
poll status bits returned by polled instruments, and for transmitting secondary
addresses following the talk or listen primary address of a device.

PPE

Defines parallel poll byte

Parallel Poll Enable
PPD

Clears parallel poll byte

Parallel Poll Disable
SCG

For transfer of a secondary address after TAG or LAG

Sec. Comm. Group

The IEEE-488 bus interface H088

The test instruments of the HAMEG Series HM8100 have all been designed to permit
their use in automated testing environments. In order to connect them to an IEEE488
bus, the interface HO88 (option) is required. Series HM8100 instruments equipped
with the IEEE488 bus interface comply with the stipulations of the IEC-625-1 and
IEEE-488 standards. If the HO88 interface is ordered together with the a HM 81 ..
instrument, it is installed within it at the factory. The HO 88 interface is also available
as a separate option for retrofitting purposes at a later time.

Software service

For operation of the Series 8100 instruments, HAMEG will be issuing software at
irregular intervals. This software will be supplied free of charge-to the owners of these
instruments. In addition, updates of the firmware contained in the instruments will be
provided in the form of new EPROMs at cost price. In order to- benefit from this
software service, it is sufficient to register by sending in your name and address and
the serial number of-your HAMEG instrument.

Installation

An HM 81 .. instrument can be easily and unproblematically retrofitted by the user
with the H088 Interface. The first step is to remove the instrument enclosure. To do

Содержание HO88

Страница 1: ...Instruments HO88 DEUTSCH ENGLISH FRAN AIS ESPA OL MANUAL HANDBUCH MANUEL...

Страница 2: ...Deutsch 3 English 13 Fran ais 22 Espa ol 28 St 1001 Zim tke Printed in Germany...

Страница 3: ...Me ger te Verwendung findet F higkeiten des IEEE Bus Zum Betrieb mehrerer Ger te an einer Schnittstelle erh lt jedes Ger t eine nur von ihm benutzte Adresse im Zahlenbereich 0 bis 30 Die Adresse wird...

Страница 4: ...le zur Kommunikation mit Peripherieger ten ist vom Ablauf der Bussteuerung bis zur Steckerbelegung so exakt definiert wie der IEEE Bus F r die Verbindung von Ger ten untereinander bestehen bis auf die...

Страница 5: ...inheitlich von allen Me ger te und Computerherstellern benutzt IEEE Bus und Computer verwenden somit den gleichen Zeichensatz In der Regel m ssen Zeichens tze nicht mehr zwischen Rechner und IEEE Bus...

Страница 6: ...N logisch 0 5V oder als Befehle ATN logisch 1 0V zu interpretieren sind Die bergabe der Daten erfolgt mit DAV NRFD und NDAC IFC IFC InterFace Clear kann immer nur vom steuernden Controller gesetzt wer...

Страница 7: ...werden mit Hilfe der bergabeleitung DAV NRFD und NDAC bertragen Ein festgelegtes Endezeichen und oder EOI kennzeichnen das Ende der bertragung Ger tenachrichten bestehen immer aus einem ger tespezifis...

Страница 8: ...gen Parallel Poll Enable PPD Parallel Poll Byte l schen Parallel Poll Disable SCG bertragen der Sekund radresse nach TAG oder LAG Secondary Command Group Die IEEE Bus Schnittstelle HO88 Allgemeines Di...

Страница 9: ...wird empfohlen die Originalverpackung aufzubewahren Transportsch den sind vom Garantieanspruch ausgeschlossen Einbau Der nachtr gliche Einbau der Schnittstelle HO88 ist sehr einfach und ohne Probleme...

Страница 10: ...nschaften Nicht alle Nachrichten m ssen von jedem Ger t decodiert werden Schnittstellennachrichten werden nur dann decodiert wenn das Ger t die hierzu notwendige Ausr stung besitzt Ein Ger t das nur L...

Страница 11: ...chung bei Adressierung als H rer Listener H rfunktion L4 Basic Listener Automatische Adressenl schung bei Adressierung als Sprecher Talker Handshake Sender und Empf nger SH1 AH1 SH1 und AH1 bedeuten g...

Страница 12: ...s ein Ger te oder auch Interfacebefehl gemischt enthalten sein die dann nacheinander abgearbeitet werden Diese m ssen lediglich durch die oben genannten Zeichen voneinander getrennt sein Eingangsdelim...

Страница 13: ...rm IEEE 488 bus will be used since the 24 pin connector specified by the American standard is used by HAMEG s test instruments Capabilities of the IEEE 488 bus For operation of a number of devices con...

Страница 14: ...ith 3 mA each 48 mA The IEEE 488 bus uses negative logic with standard TTL logic levels between 0V and 5V Logic 0 is a TTL high level 2 0V and logic 1 is a TTL low level 0 8V Figure 1 IEEE 488 bus con...

Страница 15: ...OUT is used for transfer of one bit of each 8 bit data word Most data use the 7bit ASCII or ISO code set Each letter or character is represented as a 7 bit word Figure 4 The 8th bit DIO8 is either un...

Страница 16: ...88 bus They can be influenced by the user ATN IFC and REN can only be driven by the System Controller SRQ can only be driven by a peripheral device EOI can be driven by either the CIC or a peripheral...

Страница 17: ...ng transfer of this type of message The data are transferred with aid of the handshake lines DAV NRFD and NDAC The end of a transmission is identified by a line feed and or asserting EOI Device depend...

Страница 18: ...el Poll Disable SCG For transfer of a secondary address after TAG or LAG Sec Comm Group The IEEE 488 bus interface H088 The test instruments of the HAMEG Series HM8100 have all been designed to permit...

Страница 19: ...ts connected to an IEEE488 bus must receive unique device addresses This is done using the 5 DIP switches to the left of the IEEE488 bus connector on the instruments rear panel The switches are binary...

Страница 20: ...allowed to generate the handshakecycle for receiving data Talker T T0 1 Capabilities for transmission of data Extended Talker Like Talker but TE TE0 8 using secondary addresses as well Listener Capabi...

Страница 21: ...ce Clear DC1 The HM81 can be initialized to a predefined cleared state either selectively or together with all other instruments on the bus Device Trigger DT1 A new measurement can be triggered by rem...

Страница 22: ...ersement des adaptateurs appropri s sont disponibles Dans la description ci apr s il ne sera utilis que le terme Bus IEEE les appareils de mesure HAMEG tant quip s de connecteurs 24 bro ches qui rel v...

Страница 23: ...2 0V et un 1 logique une tension 0 8V Fig 1 Assignation des broches pour bus IEEE Attention le connecteur 25 broches IEC 625 peut facilement tre confondu avec celui usuel et galement de 25 broches pr...

Страница 24: ...Fig 4 Le bit de poids le plus fort DI08 n est g n ralement pas utilis dans la transmission des donn es et n a donc aucune signi fication Dans la pratique les diff rentes combinaisons de bits sont repr...

Страница 25: ...Si ATN IFC et REN ne peuvent tre utilis s que par le contr leur c est dire I ordinateur SRQ ne peut tre contr l que par une unit p riph rique et seul le circuit EOI peut tre op r aussi bien par le co...

Страница 26: ...ommande manuelle SRQ Demande de service ATN Type de message adresse ou donn es EOI Fin de la transmission EOl et Parallel Poll Interrogation des tats ATN Messages multi fils Les messages multi fils pe...

Страница 27: ...ogation en parall le pour d sactiver I tat Parallel Poll et transmettre I adresse secondaire apr s I adressage de I metteur ou du r cepteur PPE D finition de l octet d interrogation en parall le Paral...

Страница 28: ...I est pourtant possible que la d faillance d un composant n intervienne qu apr s un service prolong C est pourquoi HAMEG accorde tous ses produits une garantie de fonctionnement de 2 ans Celle ci est...

Страница 29: ...lus communiquer avec le contr leur Propri t s de I interface Chacun des appareils n a pas besoin de d coder tous les messages les messages d interface ne sont d cod s que si I appareil adress dispose...

Страница 30: ...s instruments ou avec le contr leur Fonction metteur Talker T5 L appareil met il parle les r sultats de ses mesures vers d autres appareils ou vers un contr leur T5 signifie galement que I appareil pe...

Страница 31: ...quentiellement Lorsqu une cha ne comporte des commandes qui s excluent mutuellement seule la derni re commande sera retenue Par exemple I envoi de la cha ne FRA FRB aura pour effet de placer I appare...

Страница 32: ...norma IEC 625 existen adaptadores A continuaci n se utiiizar el t rmino bus IEEE ya que el conector de 24 terminales correspondiente a la norma IEEE 488 es ei utilizado por los instrumentos de medida...

Страница 33: ...15 aparatos se da a causa de la restringida potencia de excitaci n de los componentes de la interfaz Dicha potencia est definida como 48 mA El consumo de cada excitador es de 3 mA 15 aparatos 1 contro...

Страница 34: ...iza para transferir uno de los ocho bits de que se compone una palabra Para la transferencia se utilizan ios c digos ISO 7 Bit o ASCII de 7 bits Cada letra o car cter se representa como una palabra de...

Страница 35: ...tiene influencia sobre ellas Las lineas ATN IFC y REN s lo pueden ser manejadas des de la unidad de control ordenador El control de la I nea SRO s lo puede Ilevarse a cabo desde un aparato perif rico...

Страница 36: ...servicio ATN Mensaje dep disposit de interfaz EOI Fin del mensaje EOIy Parallel Poll exploraci n parallelo ATN Mensajes de varias lineas Esta categor a comprende tanto mensajes dependientes del dispos...

Страница 37: ...control a otro controlador Take Control Instrucciones universales no direccionadas Tienen efecto sobre todos ios aparatos LLO Imposibilita la operaci n manual Local Lock Out DCL Inicializa el instrum...

Страница 38: ...alquier tipo de fuente de tensi n Garant a Antes de salir de f brica todas las interfaces H088 pasan por un control de calidad y se someten a una prueba de quemado durante 24 horas aprox La operaci n...

Страница 39: ...l forma que sta imprima continuamente los datos del contador El modo Talk only s lo habla no debe nunca utilizarseen la configuraci n de un bus junto con un controlador ya que entonces el contador no...

Страница 40: ...receptor menos Listen only solo escucha Service Request SR1 equipo completo de instrucciones petici n de servicio Control remoto RL1 equipo completo de instrucciones manual Polling en paralelo PP0 no...

Страница 41: ...a transferencia de datos pueden utilizarse como separadores iosca racteres sigu jentes punto y coma o 38h coma o 2Ch espacio o 20h Dentro de una transferencia de datos estos caracteres permiten asimis...

Страница 42: ...chlu belegung entnehmen Sie folgender Tabelle The ribbon cable connections are as follows CON2 CON2 CON2 CON2 CON2 CON2 HM8130 CON1 HM8130 CON1 CON1 CON1 CON1 CON1 Einbauanleitung Installation Systemg...

Страница 43: ...43 Sous r serve de modifications Reservado el derecho de modificaci n Appendix B ASCII CODE CHART...

Страница 44: ...nderungen vorbehalten Subject to change without notice 44...

Страница 45: ...45 Sous r serve de modifications Reservado el derecho de modificaci n...

Страница 46: ...nderungen vorbehalten Subject to change without notice 46...

Страница 47: ......

Страница 48: ...4 1855 E mail hamegny aol com Great Britain HAMEG LTD 74 78 Collingdon Street LUTON Bedfordshire LU1 1RX Phone 01582 413174 Telefax 01582 456416 E mail sales hameg co uk Spain HAMEG S L Villarroel 172...

Отзывы: