ORCA-Lightning Digital CMOS Camera C14120-20P Instruction manual_Ver.2.5
51
13-3 AT WATER-COOLING
Symptom
Cause
Measures
Chapter
Cooling or circulation
does not start
The power supply of circulating
water cooler is not turning on
Turn on the power supply of the
circulating water cooler.
The cooling method of DCAM
Configurator is set to “Air-cooling”
Change the setting to
“Water-cooling”.
Quantity of water is insufficient
Supply sufficient quantity of
water (Refer to the instruction
manual of the circulating water
cooler).
Noise appears on the
screen
The temperature of cooling water is
high (more than 25 °C)
Lower the temperature to 25 °C.
The flow of cooling
water is unnatural
The inside of the hose is clogged up
or is dirty
Wash dirt away by water current
or water pressure. If the
symptom is unimproved,
exchange it for a new hose.
Water leak
The connection of the hose is loose
Reconnect the hose.
The diameter of the hose does not fit
the connector of the circulating water
cooler
Exchange it for the hose in
accord with the diameter of the
connector.
The hose is damaged
Exchange it for a new hose.