ORCA-Lightning Digital CMOS Camera C14120-20P Instruction manual_Ver.2.5
13
f. Power switch [POWER]
The power is turned on/off.
When the camera is turned on, the camera starts to initialize and after the initialization is
complete, the camera can be operated.
g. DC power input connector [DC IN]
This is the power supply terminal. Use the accessory AC adapter.
h. WATER connector for Water-cooling [WATER]
They connect to the water connectors on the circulating water cooler with the cooling water
hoses. The connector position of WATER IN/OUT is not specified. If Water-cooling is used,
refer to 11. “WHEN USING WATER-COOLING METHOD”.
i. Fan for Air-cooling
This is a fan for heat ventilation that blown air inside the camera.
The fan does not work in Water-cooling.
•
If the camera is being operated in an enclosed environment, ensure to keep clearance at
least 10 cm from both intake and exhaust vents when setting up.
•
To prevent overheating inside the camera, do not wrap the camera in cloth or other
material, or block the camera’s ventilation.
j. Air outlet
This is the outlet for heat ventilation.
•
If the camera is being operated in an enclosed environment, ensure to keep clearance at
least 10 cm from both intake and exhaust vents when setting up.
•
To prevent overheating inside the camera, do not wrap the camera in cloth or other
material, or block the camera’s ventilation.