11. WHEN USING WATER-COOLING METHOD
Water-cooling method is used when the camera is used in a place where the air conditioning is unstable.
•
When using Water-cooling method, follow this chapter and chapter 8.
11-1 BEFORE USING THE CAMERA AT WATER-COOLING
(1) Change the cooling method by “DCAM Configurator”
The default setting of cooling method is Air-cooling. The setting is changed by the “DCAM
Configurator” software. When the setting is changed, restart the camera. By restarting, the
setting is reflected on the camera. Refer to 11-5 “SWITCHING THE COOLING METHOD OF
“DCAM CONFIGURATOR” for the setting procedure.
(2) Cooling water
a. Requirement
Hamamatsu recommends using soft water (except pure water) for cooling water.
If you plan on using water other than soft water as recommended for example antifreeze etc,
choose the water which followed notices of 12-2. “WHEN USING AT WATER-COOLING” or
contact a Hamamatsu subsidiary or your local distributor.
b. Recommendation temperature
Hamamatsu recommends 25 °C for Circulating water temperature.
For the appropriate temperature range of the cooling water, confirm with the instruction manual
of your circulating water cooler.
(3) Circulating water cooler
Handle the circulating water cooler and the cooling water according to an instruction manual of
the circulating water cooler. Proper performance may not be achievable if a non-recommended
circulating water cooler is used.
(4) Cooling water hose
a. Configurations
The hose has a blue hose and a gray hose.
[Diameter of hoses]
Internal
External
Blue hose
4 mm
6 mm
Gray hose
8 mm
13.5 mm
Blue hose
Gray hose