ORCA-Lightning Digital CMOS Camera C14120-20P Instruction manual_Ver.2.5
33
10-4 READOUT METHODS
The readout methods can set units for reading out images. In the Sub-array readout, the readout area
is also set. The methods have the below.
(
: enable,
−
: disable)
Readout methods
Unit of readout pixel
(Horizontal×Vertical)
Readout area
(Horizontal×Vertical)
Normal
Area
Mode
Lightsheet
Readout
Mode
Normal readout
1×1 (1 pixel)
Whole area: 4608 × 2592
Binning readout
2×2
Whole area: 2304 × 1296
−
4×4
Whole area: 1152 × 648
Sub-array readout
1×1 (1 pixel)
Designated area
(1) Normal readout (Full resolution readout)
Perform charge readout from camera individually for all pixels.
(2) Binning readout
With this camera, 2×2 binning and 4×4 binning readout are available by adding the signal of
adjacent pixels in the digital domain. Binning readout is a method for achieving high sensitivity in
exchange for losing resolution. The mode is not supported at Lightsheet Readout Mode.
(3) Sub-array readout
The Sub-array readout is a procedure only a region of interest is scanned. It is possible to
increase the frame rate by reducing the number of vertical lines scanned. (when using the
“DCAM-API” software.) In the readout, binning configuration is enabled.
Camera readout modes
Specified
Unit of setting
Horizontal
Vertical
Normal Area Mode
Position (Start position)
4 pixels of intervals
4 pixels of units
by 4 lines
Area size
by 128 pixels
by 4 lines
Lightsheet Readout Mode
Position (Start position)
by 128 pixels
by 4 lines
Area size
by 128 pixels
by 4 lines