ORCA-Lightning Digital CMOS Camera C14120-20P Instruction manual_Ver.2.5
36
(2) Programmable timing output
a. In Normal Area Mode
By using the programmable timing output, synchronizing external equipment is simple. A system
that needs simple timing signal does not require a delay unit or pulse generator. It is possible to
program and output a pulse that has an optional pulse width and an optional delay time to Read
End (the end of readout timing) or Vsync. The setting range for delay time is 0 µs to 10 s, and the
setting range for pulse width is 1 µs to 10 s.
The relation between the parameter which can be set with each reference signal and an output
signal becomes below.
Reference signal Output signal
Read End
Camera outputs a pulse after certain delay, from the end of sensor readout. Also the
pulse width can be set.
Vsync
Camera outputs a pulse after certain delay, from the beginning of readout. Also the
pulse width can be set.
Figure 10-13 (Example for Programmable timing output in Normal Area Mode)