If the hose size on circulating water cooler is the same as blue hose, remove gray hose from
the joint part. The gray hose can be removed when blue hose is pulled with pushing the button
of the joint on gray hose.
b. Connections to connectors of the circulating water cooler
When connecting the hoses to each connector of the camera and the circulating water cooler,
handle according to 11-3 "CONNECTIONS”.
The hoses do not have to be removed from the connector every time. If disconnecting the
hoses from connectors, handle according to 11-6 "DISCONNECTION OF WATER COOLING
HOSES”.
- Stop water circulation when connecting / disconnecting the hose, and turn off the power of the
camera and the circulating water cooler.
- Confirm that cooling water stops.
- Prepare water absorption sheet (such as Waste, Towel or so) and catch pan in order to avoid
water drop or water splash.
c. Deterioration of hoses
Replace the water hose with a new one whenever it cannot keep 0.45 L/min flow rate for water
circulation due to the hose deterioration.
(5) Cautions during Water-cooling
Be always careful about the following during Water-cooling.
- Keep 0.45 L/min flow rate for water circulation.
- Check cooling water is circulating properly and the camera is cooled properly.
- Do not stop the cooling while the power supply is supplied to the camera.
- Be careful not to be scattered water drop or water splash around peripherals. Prepare water
absorption sheet (such as Waste, Towel or so) and catch pan in order to avoid them.
11-2 PREPARING
preparation
Hamamatsu option
Circulating water cooler
C3142-11
Cooling water hose (2 hoses)
A10788-04
Cooling water
―
•
When you need our option the circulating water cooler and the cooling water hoses , please contact a
Hamamatsu subsidiary or your local distributor.
•
Handle the circulating water cooler and the cooling water according to this manual and the instruction
manual of the circulating water cooler.
Joint
Button