background image

Tafonius

 Auxiliary Mode Operating Manual  

Printed: 28/Feb/11 

 

Page 8 of 15 

DOCA4956 

Initialising the Piston 

As stated, before Tafonius can be used the system needs to locate the zero position, the 
bottom of travel, of the piston. 

The RUN button will be flashing, after the boot sequence, while waiting for the user.  Make 
sure there is nothing attached to the Y-piece before pushing RUN the button.  Once the 
button is pushed the screen shows the message 

MOVING PISTON TO ZERO.  

During this 

piston movement the airway pressure is monitored.  If there is a patient attached or an 
obstruction to airflow then the following message appears: 

ERROR. PRESSURE TOO HIGH  
DISCONNECT PATIENT 
THEN PUSH RUN BUTTON 

 
Clear the obstruction or remove the patient before continuing.  There are 2 other possible 
reasons why the initialisation procedure may be interrupted:  The first is that the cylinder is 
not closed properly.  The second is that the vacuum surrounding the piston is too low.  In 
either of these instances warning messages will appear alerting you to the cause and 
remedy of the interruption.  Once the piston reaches the bottom of its travel, a zero reading 
is taken and 

ZEROING PISTON POSITON

 is displayed briefly.  Three beeps are heard and 

the control screen is displayed, see below.  The piston is at the very bottom waiting for the 
user to pre-fill the system with oxygen and anaesthetic for induction.  

Tafonius

  is ready 

for use in its Auxiliary Mode. 

 

The Control Screen 

Before continuing here is an explanation of what you see on the auxiliary screen. It is 
divided into 4 main regions: 

 
1) Status and message information 

Area 1 shows status information on the system batteries, whether the machine is running on 
mains or battery power, the software version, the vacuum level around the piston and also 

 

1)  Status and message information 
2)  Piston Graphic and information  
3)  Resultant Ventilation values 
4)  Patient Ventilator settings & 

Measured values 

Содержание Vetronic Tafonius

Страница 1: ...Tafonius Auxiliary Mode Operating Manual A USERS GUIDE v09Feb2011...

Страница 2: ...Tafonius Auxiliary Mode Operating Manual Printed 28 Feb 11 Page 2 of 15 DOCA4956 Notes...

Страница 3: ...ator settings recorded values 10 Control of the Ventilator in Auxiliary Mode 10 Setting Values using the control knobs Change Commit 10 Control setting validation 10 THE CONTROL KNOBS 10 Tidal Volume...

Страница 4: ...A4956 PISTON POSITION INFO ERROR 13 Alarms 14 LOW BREATHING SYSTEM PRESSURE 14 OXYGEN SUPPLY PRESSURE LOW 14 BATTERY 1 CRITICALLY LOW and or 14 BATTERY 2 CRITICALLY LOW 14 ENCODER BOARD HAS FAILED 14...

Страница 5: ...re will very rarely be needed or used If there is a discrepancy between the two offset values seen on the screen below it usually indicates some problem other that pressure sensor drift Most frequentl...

Страница 6: ...the individual battery voltages are measured and its value displayed on the screen e g CHECKING DRIVE VOLTAGE 27 9 14 0 13 9 If the supply voltage 27 9 above is less than 16 0 volts then the shutdown...

Страница 7: ...d device If no USB device is detected as is the case when running without a PC monitor the auxiliary controller will change the display to below left and wait for the Tafonius software to boot up and...

Страница 8: ...nder is not closed properly The second is that the vacuum surrounding the piston is too low In either of these instances warning messages will appear alerting you to the cause and remedy of the interr...

Страница 9: ...he level of inspiratory flow in liters per minute the patient must generate or exceed to trigger and assisted breath In STANDBY mode In STANDBY spontaneous breathing mode this effort level can be set...

Страница 10: ...control knob then after 4 seconds the setting will revert to the original value This process prevents accidental or inadvertent changing of ventilation parameters Control setting validation There are...

Страница 11: ...essure deviation in excess of 0 5 cmH2O from the ambient pressure is detected the piston is moved so as to negate this pressure change If the patient breathes out then the piston withdraws effectively...

Страница 12: ...respectively The tidal volume setting is then increased to 4 0 This increases the upper limit to 6 5L so the piston is allowed to rise to that level by virtue of the continuously flowing fresh gas flo...

Страница 13: ...t will return MAX PRESSURE EXCEEDED LIMIT The airway pressure exceeded the setting of the Maximum Working Pressure Limit The Alert will stay on screen until cleared manually LOW SYSTEM RESERVE VOLUME...

Страница 14: ...linders has fallen below 40 psi This alarm can only be cleared by connecting a source of oxygen with a pressure greater than 40psi BATTERY 1 CRITICALLY LOW and or BATTERY 2 CRITICALLY LOW Either one o...

Страница 15: ...ght 140kg when fully fitted no oxygen cylinders Construction Powder coated stainless steel and aluminum ELECTRICAL Power Input 90 240v AC 50 60Hz Universal input Power consumption 400W max Operational...

Отзывы: