Hallowell EMC Vetronic Tafonius Скачать руководство пользователя страница 9

Tafonius

 Auxiliary Mode Operating Manual  

Printed: 28/Feb/11 

 

Page 9 of 15 

DOCA4956 

displays any error or warning messages.  The battery information shows the actual voltages 
of the two batteries used in 

Tafonius

 and whether or not they are being charged  When 

there are no warning or error messages, the name 

TAFONIUS

 is displayed. 

2) Piston Graphic and Information 

Area 2 shows a representation of the piston.  The dark area shows the volume of gas in the 
piston and the numerical value at the top of the list to the right gives the amount in litres. 
See the photo on the front cover as the piston in the photo above is at the bottom with 0.0 
litres under it.  The piston position and volume is constantly updated and shows the position 
of the piston at all times.  Two asterix (* *) will appear at the top of the piston area whenever 
the Dump Valve, an electric pop-off valve, is open.  

Below the volume display in litres are three settings that are not otherwise always display on 
the screen.  

MWPL 

The Maximum Working Pressure Limit in cmH

2

O is generally seen above the right most 

control knob used to set its value however it will not be displayed there when using the 
Assist Mode.  The MWPL is set as a high pressure safety limit the value of which is 
continuously compared to the breathing system pressure.  

Tafonius

 is a volume or time 

cycled ventilator so the MWPL does NOT set how high the pressure will go but if the 
pressure should reach the set level the machine will alarm and immediately, terminate 
inspiration and switch to the expiratory phase of the breathing cycle. 

ASSIST 

The ASSIST function is a mechanism that detects a patient’s efforts to breathe then gives 
an assisted breath if that effort meets or exceeds a user set minimum level.  The value next 
to the text 

ASST 

is the level of inspiratory flow, in liters per minute, the patient must 

generate or exceed to trigger and assisted breath. 

In STANDBY mode 

In STANDBY, spontaneous breathing, mode this effort level can be set but is not active until 
the IPPV/RUN button is pushed.  To use the ASSIST function, press the ASSIST button 
once, the ASSIST button will flash slowly and the MWPL legend on the auxiliary screen will 
be replaced with the text “SENS” for sensitivity followed by a value e.g. 60 LPM. This value 
of 60 litres per minute is the inspiratory flow rate that the patient must meet or exceed in 
order to trigger the next IPPV stroke.  This will not happen unless the ventilator is in IPPV 
mode.  Pressing the RUN/IPPV button will cause the IPPV button to be lit and the ASSIST 
button will then remain lit continuously. 

In IPPV mode 

Alternatively the ASSIST mode can be chosen when the patient is receiving IPPV.  In this 
case the ASSIST mode will be lit continuously to show that the ASSIST mode is active.  In 
this state the ventilator is waiting for an effort from the patient.  If this effort level is not met 
within the time required for 3 breath cycles based on the set RR a mandatory breath is 
given.  If the patient triggers a breath then the timing of the mandatory breath is reset.  In 
either case, the delivered breath will be based on the set TV and I-time.  Pressing the 
RUN/IPPV button will return the ventilator to spontaneous mode and cancel the ASSIST 
function. 

This is a new feature and we currently have no guidelines on values to set.  The software 
offers a range of between 1 LPM and 200 LPM as trigger values.  We would welcome 
feedback on the settings found useful or normal in practice.  

3) Resultant Ventilation values 

Area 3 shows the values that result from the settings of the independent variables Tidal 
Volume, Respiratory Rate and Inspiratory Time.  The dependent variables displayed in this 
area are Minute Volume (MV) in liters per minute (lpm), Inspiratory Flow Rate (IF) in lpm, I:E 

Содержание Vetronic Tafonius

Страница 1: ...Tafonius Auxiliary Mode Operating Manual A USERS GUIDE v09Feb2011...

Страница 2: ...Tafonius Auxiliary Mode Operating Manual Printed 28 Feb 11 Page 2 of 15 DOCA4956 Notes...

Страница 3: ...ator settings recorded values 10 Control of the Ventilator in Auxiliary Mode 10 Setting Values using the control knobs Change Commit 10 Control setting validation 10 THE CONTROL KNOBS 10 Tidal Volume...

Страница 4: ...A4956 PISTON POSITION INFO ERROR 13 Alarms 14 LOW BREATHING SYSTEM PRESSURE 14 OXYGEN SUPPLY PRESSURE LOW 14 BATTERY 1 CRITICALLY LOW and or 14 BATTERY 2 CRITICALLY LOW 14 ENCODER BOARD HAS FAILED 14...

Страница 5: ...re will very rarely be needed or used If there is a discrepancy between the two offset values seen on the screen below it usually indicates some problem other that pressure sensor drift Most frequentl...

Страница 6: ...the individual battery voltages are measured and its value displayed on the screen e g CHECKING DRIVE VOLTAGE 27 9 14 0 13 9 If the supply voltage 27 9 above is less than 16 0 volts then the shutdown...

Страница 7: ...d device If no USB device is detected as is the case when running without a PC monitor the auxiliary controller will change the display to below left and wait for the Tafonius software to boot up and...

Страница 8: ...nder is not closed properly The second is that the vacuum surrounding the piston is too low In either of these instances warning messages will appear alerting you to the cause and remedy of the interr...

Страница 9: ...he level of inspiratory flow in liters per minute the patient must generate or exceed to trigger and assisted breath In STANDBY mode In STANDBY spontaneous breathing mode this effort level can be set...

Страница 10: ...control knob then after 4 seconds the setting will revert to the original value This process prevents accidental or inadvertent changing of ventilation parameters Control setting validation There are...

Страница 11: ...essure deviation in excess of 0 5 cmH2O from the ambient pressure is detected the piston is moved so as to negate this pressure change If the patient breathes out then the piston withdraws effectively...

Страница 12: ...respectively The tidal volume setting is then increased to 4 0 This increases the upper limit to 6 5L so the piston is allowed to rise to that level by virtue of the continuously flowing fresh gas flo...

Страница 13: ...t will return MAX PRESSURE EXCEEDED LIMIT The airway pressure exceeded the setting of the Maximum Working Pressure Limit The Alert will stay on screen until cleared manually LOW SYSTEM RESERVE VOLUME...

Страница 14: ...linders has fallen below 40 psi This alarm can only be cleared by connecting a source of oxygen with a pressure greater than 40psi BATTERY 1 CRITICALLY LOW and or BATTERY 2 CRITICALLY LOW Either one o...

Страница 15: ...ght 140kg when fully fitted no oxygen cylinders Construction Powder coated stainless steel and aluminum ELECTRICAL Power Input 90 240v AC 50 60Hz Universal input Power consumption 400W max Operational...

Отзывы: