Tafonius

 Auxiliary Mode Operating Manual  

Printed: 28/Feb/11 

 

Page 6 of 15 

DOCA4956 

 

Should one chooses to enter the Pressure Sensor Setup routine there are two options: 

  Pressing any of the four control knobs will use the factory default value as shown. 

  Pressing the RUN / IPPV button will use the actual current measured value, which is 

constantly being updated on the screen. 

The chosen value is stored in the non-volatile memory and will remain unchanged until the 
Pressure Sensor Setup routine is run again. 

Use the Pressure Sensor setup routine if there is piston movement when no patient is 
attached the wye piece is not capped AND you have flushed the sensing line. 

The initialisation procedure continues into the Auxiliary Controller Boot Up Sequence. 

Auxiliary Controller Boot Up Sequence 

After the option for pressure sensor setup, the following sequence occurs: 

INITIALISING DRIVE CONTROL 

At this stage the servomotor drive, the ViX Drive, is presumed un-initialised and its 
initialisation begins.  The software version is displayed at the bottom left hand corner of the 
screen.  The 24 volt supply line feeding the motor drive as well as the individual battery 
voltages are measured and its value displayed on the screen e.g. 

 

CHECKING DRIVE VOLTAGE: 

27.9  14.0  13.9 

 
If the supply voltage, 27.9 above, is less than 16.0 volts, then the shutdown procedure starts 
as there is insufficient battery power to safely run the system.  In this instance you should 
connect the mains power and re-start the system.  

If the voltages are sufficient one beep is heard as the screen displays the following. 

VOLTAGES ALL OK 
APPLYING POWER TO VIX DRIVE: 

 
The power to the servomotor drive is applied and tested.  Then after successful 
communications have been established with the ViX drive, two beeps are heard and the 
following is displayed.  

SETTING COMMUNICATIONS TO DRIVE 
SETTING COMS RATE TO 19200 
   COMMS  -  SET AT 19200 BAUD  -  OK 

 
The servomotor drive is then initialised.  The next stage checks the Motor Drive Integrity as 
the following is displayed 

NOW TESTING VIX DRIVE INTEGRITY  

 
Assuming all is well then the message  

INITIAL DRIVE CHECK OK  

is displayed, followed by a string of ones and zeros on the screen. if at any time during this 
Auxiliary initialisation procedure an error is detected then an error message is displayed and 
the whole system shuts down.  If everything is as it should be then the screen shows  
 

INITIALISATION COMPLETE 

 
At this stage the Auxiliary Screen looks like the screen below. 
 

Содержание Vetronic Tafonius

Страница 1: ...Tafonius Auxiliary Mode Operating Manual A USERS GUIDE v09Feb2011...

Страница 2: ...Tafonius Auxiliary Mode Operating Manual Printed 28 Feb 11 Page 2 of 15 DOCA4956 Notes...

Страница 3: ...ator settings recorded values 10 Control of the Ventilator in Auxiliary Mode 10 Setting Values using the control knobs Change Commit 10 Control setting validation 10 THE CONTROL KNOBS 10 Tidal Volume...

Страница 4: ...A4956 PISTON POSITION INFO ERROR 13 Alarms 14 LOW BREATHING SYSTEM PRESSURE 14 OXYGEN SUPPLY PRESSURE LOW 14 BATTERY 1 CRITICALLY LOW and or 14 BATTERY 2 CRITICALLY LOW 14 ENCODER BOARD HAS FAILED 14...

Страница 5: ...re will very rarely be needed or used If there is a discrepancy between the two offset values seen on the screen below it usually indicates some problem other that pressure sensor drift Most frequentl...

Страница 6: ...the individual battery voltages are measured and its value displayed on the screen e g CHECKING DRIVE VOLTAGE 27 9 14 0 13 9 If the supply voltage 27 9 above is less than 16 0 volts then the shutdown...

Страница 7: ...d device If no USB device is detected as is the case when running without a PC monitor the auxiliary controller will change the display to below left and wait for the Tafonius software to boot up and...

Страница 8: ...nder is not closed properly The second is that the vacuum surrounding the piston is too low In either of these instances warning messages will appear alerting you to the cause and remedy of the interr...

Страница 9: ...he level of inspiratory flow in liters per minute the patient must generate or exceed to trigger and assisted breath In STANDBY mode In STANDBY spontaneous breathing mode this effort level can be set...

Страница 10: ...control knob then after 4 seconds the setting will revert to the original value This process prevents accidental or inadvertent changing of ventilation parameters Control setting validation There are...

Страница 11: ...essure deviation in excess of 0 5 cmH2O from the ambient pressure is detected the piston is moved so as to negate this pressure change If the patient breathes out then the piston withdraws effectively...

Страница 12: ...respectively The tidal volume setting is then increased to 4 0 This increases the upper limit to 6 5L so the piston is allowed to rise to that level by virtue of the continuously flowing fresh gas flo...

Страница 13: ...t will return MAX PRESSURE EXCEEDED LIMIT The airway pressure exceeded the setting of the Maximum Working Pressure Limit The Alert will stay on screen until cleared manually LOW SYSTEM RESERVE VOLUME...

Страница 14: ...linders has fallen below 40 psi This alarm can only be cleared by connecting a source of oxygen with a pressure greater than 40psi BATTERY 1 CRITICALLY LOW and or BATTERY 2 CRITICALLY LOW Either one o...

Страница 15: ...ght 140kg when fully fitted no oxygen cylinders Construction Powder coated stainless steel and aluminum ELECTRICAL Power Input 90 240v AC 50 60Hz Universal input Power consumption 400W max Operational...

Отзывы: