Hallowell EMC Vetronic Tafonius Скачать руководство пользователя страница 5

Tafonius

 Auxiliary Mode Operating Manual  

Printed: 28/Feb/11 

 

Page 5 of 15 

DOCA4956 

Running 

Tafonius

 with the Auxiliary Control System 

Turning 

T afonius

 On 

Press the soft ON/OFF control button (1).  The auxiliary screen (2) and the soft

 

ON/OFF 

button immediately illuminate.  There is a single output (beep) from the speaker and the 
auxiliary screen shows “PUSH ANY KNOB FOR PRESSURE SENSOR SETUP”.  The user 

has 2 seconds to push any of the rotary knobs to enter the Pressure Sensor setup screen.  
Experience over the last 4 years has indicated that this feature will very rarely be needed or 
used.  If there is a discrepancy between the two offset values seen on the screen below, it 
usually indicates some problem other that pressure sensor drift.  Most frequently you’ll find 
water in the pressure sampling line.  It is recommended that you  blow out the pressure 
sensing line on a regular basis.  To facilitate this process all machines produced after T21 
will be shipped with a stopcock at the machine end of the sampling line.  You can easily add 
one to older machines yourself.  To flush the line out attach a 60cc syringe to the open port 
on the stopcock, turn the stopcock handle to shut off the port 

entering

 the machine and 

flush the line with the air in the 60cc syringe as shown to the right below. 

 

 

 

 

 

Содержание Vetronic Tafonius

Страница 1: ...Tafonius Auxiliary Mode Operating Manual A USERS GUIDE v09Feb2011...

Страница 2: ...Tafonius Auxiliary Mode Operating Manual Printed 28 Feb 11 Page 2 of 15 DOCA4956 Notes...

Страница 3: ...ator settings recorded values 10 Control of the Ventilator in Auxiliary Mode 10 Setting Values using the control knobs Change Commit 10 Control setting validation 10 THE CONTROL KNOBS 10 Tidal Volume...

Страница 4: ...A4956 PISTON POSITION INFO ERROR 13 Alarms 14 LOW BREATHING SYSTEM PRESSURE 14 OXYGEN SUPPLY PRESSURE LOW 14 BATTERY 1 CRITICALLY LOW and or 14 BATTERY 2 CRITICALLY LOW 14 ENCODER BOARD HAS FAILED 14...

Страница 5: ...re will very rarely be needed or used If there is a discrepancy between the two offset values seen on the screen below it usually indicates some problem other that pressure sensor drift Most frequentl...

Страница 6: ...the individual battery voltages are measured and its value displayed on the screen e g CHECKING DRIVE VOLTAGE 27 9 14 0 13 9 If the supply voltage 27 9 above is less than 16 0 volts then the shutdown...

Страница 7: ...d device If no USB device is detected as is the case when running without a PC monitor the auxiliary controller will change the display to below left and wait for the Tafonius software to boot up and...

Страница 8: ...nder is not closed properly The second is that the vacuum surrounding the piston is too low In either of these instances warning messages will appear alerting you to the cause and remedy of the interr...

Страница 9: ...he level of inspiratory flow in liters per minute the patient must generate or exceed to trigger and assisted breath In STANDBY mode In STANDBY spontaneous breathing mode this effort level can be set...

Страница 10: ...control knob then after 4 seconds the setting will revert to the original value This process prevents accidental or inadvertent changing of ventilation parameters Control setting validation There are...

Страница 11: ...essure deviation in excess of 0 5 cmH2O from the ambient pressure is detected the piston is moved so as to negate this pressure change If the patient breathes out then the piston withdraws effectively...

Страница 12: ...respectively The tidal volume setting is then increased to 4 0 This increases the upper limit to 6 5L so the piston is allowed to rise to that level by virtue of the continuously flowing fresh gas flo...

Страница 13: ...t will return MAX PRESSURE EXCEEDED LIMIT The airway pressure exceeded the setting of the Maximum Working Pressure Limit The Alert will stay on screen until cleared manually LOW SYSTEM RESERVE VOLUME...

Страница 14: ...linders has fallen below 40 psi This alarm can only be cleared by connecting a source of oxygen with a pressure greater than 40psi BATTERY 1 CRITICALLY LOW and or BATTERY 2 CRITICALLY LOW Either one o...

Страница 15: ...ght 140kg when fully fitted no oxygen cylinders Construction Powder coated stainless steel and aluminum ELECTRICAL Power Input 90 240v AC 50 60Hz Universal input Power consumption 400W max Operational...

Отзывы: