The auxiliary controller next displays
CHECKING FOR PC CONNECTION
and waits while it checks for the presence of a USB attached device. If no USB device is
detected, as is the case when running without a PC monitor the auxiliary controller will
change the display (to below left) and wait for the
Tafonius
software to boot up and take
over control. Should the
Tafonius
software not claim control within two minutes
controller proceeds with piston initialisation and displays instructions to
DISCONNECT THE
PATIENT AND PUSH THE RUN BUTTON
.