Hallowell EMC Vetronic Tafonius Скачать руководство пользователя страница 14

Tafonius

 Auxiliary Mode Operating Manual  

Printed: 28/Feb/11 

 

Page 14 of 15 

DOCA4956 

Alarms 

These messages are accompanied by a repeating intermittent beep of approximately 1 
second duration.  A flashing text message will appear on the screen in area 1, where the 

Tafonius

 logo normally sits.  Alarms are critical events that have serious implications for 

the efficient ventilation or support of the patient. Each alarm that is shown on the screen will 
have an asterisk at each end of the text, e.g.: 

 

* OXYGEN SUPPLY PRESSURE LOW * 

 
To clear an alarm manually press and hold any of the 4 control knobs until two beeps are 
heard.  Here is a list of the alarms. 

*  LOW BREATHING SYSTEM PRESSURE  * 

This Alarm occurs when, at the end of an inspiratory phase during IPPV or Assist, a 
maximum pressure of 2cm H2O has not been reached.  This can occur when the patient 
has become disconnected from the system.  It can also occur when the patient takes a 
breath at the same time as the ventilator delivers a breath. 

The Alarm will continue until either a pressure in excess of 2cm H

2

O is registered during an 

inspiratory phase OR the alarm is cleared manually.

 

*  OXYGEN SUPPLY PRESSURE LOW  * 

The piped oxygen pressure, whether it is from the hospital supply Line, or from the E-
cylinders has fallen below 40 psi.  This alarm can only be cleared by connecting a source of 
oxygen with a pressure greater than 40psi.  

*  BATTERY 1 CRITICALLY LOW  * 

and/or

 

*  BATTERY 2 CRITICALLY LOW  * 

Either one of the batteries voltage has reached the critical level.  Below this level ventilator 
function may be severely compromised, particularly with regard to piston movement and the 
action of the Dump Valve.  

The ventilator must be connected to the mains supply 

immediately

.

 

*  ENCODER BOARD HAS FAILED  * 

The circuit board for the front control knobs has failed to respond to the main controller. It 
will not be possible to change any of the ventilation settings. 

This is a critical error and the 

patient should be changed to manual ventilation

.

 

*  CHECK POSITION SENSORS  * 

The lower or upper position sensor has been moved or is faulty. Check the alignment marks 
of the position sensors on the piston actuator. Without a lower position sensor the ventilator 
cannot be initialised and will not run. 

The patient should be changed to manual 

ventilation

.

 

*  RD VACUUM FAILURE  * 

The vacuum level in the double diaphragm has fallen below a critical level. Continued use of 
the ventilator will result in severe damage to the piston assembly. The vacuum level must be 
restored before ventilation is re-enabled. Check the connections of the hose at the vacuum 
port on the ventilator cylinder

. Until the vacuum has been restored the patient should be 

changed to manual ventilation

Содержание Vetronic Tafonius

Страница 1: ...Tafonius Auxiliary Mode Operating Manual A USERS GUIDE v09Feb2011...

Страница 2: ...Tafonius Auxiliary Mode Operating Manual Printed 28 Feb 11 Page 2 of 15 DOCA4956 Notes...

Страница 3: ...ator settings recorded values 10 Control of the Ventilator in Auxiliary Mode 10 Setting Values using the control knobs Change Commit 10 Control setting validation 10 THE CONTROL KNOBS 10 Tidal Volume...

Страница 4: ...A4956 PISTON POSITION INFO ERROR 13 Alarms 14 LOW BREATHING SYSTEM PRESSURE 14 OXYGEN SUPPLY PRESSURE LOW 14 BATTERY 1 CRITICALLY LOW and or 14 BATTERY 2 CRITICALLY LOW 14 ENCODER BOARD HAS FAILED 14...

Страница 5: ...re will very rarely be needed or used If there is a discrepancy between the two offset values seen on the screen below it usually indicates some problem other that pressure sensor drift Most frequentl...

Страница 6: ...the individual battery voltages are measured and its value displayed on the screen e g CHECKING DRIVE VOLTAGE 27 9 14 0 13 9 If the supply voltage 27 9 above is less than 16 0 volts then the shutdown...

Страница 7: ...d device If no USB device is detected as is the case when running without a PC monitor the auxiliary controller will change the display to below left and wait for the Tafonius software to boot up and...

Страница 8: ...nder is not closed properly The second is that the vacuum surrounding the piston is too low In either of these instances warning messages will appear alerting you to the cause and remedy of the interr...

Страница 9: ...he level of inspiratory flow in liters per minute the patient must generate or exceed to trigger and assisted breath In STANDBY mode In STANDBY spontaneous breathing mode this effort level can be set...

Страница 10: ...control knob then after 4 seconds the setting will revert to the original value This process prevents accidental or inadvertent changing of ventilation parameters Control setting validation There are...

Страница 11: ...essure deviation in excess of 0 5 cmH2O from the ambient pressure is detected the piston is moved so as to negate this pressure change If the patient breathes out then the piston withdraws effectively...

Страница 12: ...respectively The tidal volume setting is then increased to 4 0 This increases the upper limit to 6 5L so the piston is allowed to rise to that level by virtue of the continuously flowing fresh gas flo...

Страница 13: ...t will return MAX PRESSURE EXCEEDED LIMIT The airway pressure exceeded the setting of the Maximum Working Pressure Limit The Alert will stay on screen until cleared manually LOW SYSTEM RESERVE VOLUME...

Страница 14: ...linders has fallen below 40 psi This alarm can only be cleared by connecting a source of oxygen with a pressure greater than 40psi BATTERY 1 CRITICALLY LOW and or BATTERY 2 CRITICALLY LOW Either one o...

Страница 15: ...ght 140kg when fully fitted no oxygen cylinders Construction Powder coated stainless steel and aluminum ELECTRICAL Power Input 90 240v AC 50 60Hz Universal input Power consumption 400W max Operational...

Отзывы: