Clamping force tester TESTit – Trouble shooting
9.2 Trouble shooting table
Fault
Possible cause
Fault correction
Corrected by
After pressing the
ON button on the
measuring head
there is nothing
displayed.
The battery capacity of
the measuring head is
too low.
Charge the battery of the
measuring head.
Specialist
The ON button wasn't
pressed properly.
Press the ON button again. Specialist
There are no data to
save
The measuring head is
not turned on.
Switch on the measuring
head.
Specialist
The battery capacity of
the measuring head is
too low.
Charge the battery of the
measuring head.
Specialist
Bluetooth connection is
faulty
Check the connection of
the measuring head and
tablet.
Specialist
The tablet does not
boot
The battery is empty
Charge the battery.
Specialist
The tablet has a defect
Send the tablet back to the
manufacturer
HAINBUCH
Measuring head is
damaged
Measuring head has
fallen / struck
A factory function testing is
necessary and possibly an
exchange of components.
HAINBUCH
The display of the
measuring head is
off
The accu of the
measuring head is empty
Connect the charging cable
and recharge the
measuring head.
Specialist
The connection to
the measuring head
can not be initialized
The measuring head is
not set on
Switch on the measuring
head and repeat the
connection
Specialist
Order H49 7144.907-333