1.
Use a clean, soft cloth to remove loose debris from the end of the sensor. Rinse the sensor with
clean, warm water.
2.
Soak the sensor for 2 to 3 minutes in a soap solution.
3.
Use a soft bristle brush to scrub the entire measuring end of the sensor.
4.
If debris remains, soak the measuring end of the sensor in a dilute acid solution such as < 5%
HCl for a maximum of 5 minutes.
5.
Rinse the sensor with water and then return to the soap solution for 2 to 3 minutes.
6.
Rinse the sensor with clean water.
Always calibrate the sensor after maintenance procedures.
Troubleshooting
Sensor diagnostic and test menu
The sensor diagnostic and test menu shows current and historical information about the instrument.
Refer to
.
To access the sensor diagnostic and test menu, push the
menu
key and select SENSOR
SETUP>DIAG/TEST.
Table 2 Sensor DIAG/TEST menu
Option
Description
MODULE INFORMATION
Shows information about the sensor module.
SENSOR INFORMATION
Shows the name and serial number that was entered
by the user.
CAL DAYS
Shows the number of days since the last calibration.
CAL HISTORY
Shows a list of all calibrations by date/time stamp. Use
the arrows keys to select a calibration and push
enter
to view the details.
RESET CAL HISTORY
Resets the calibration history for the sensor (requires
service-level passcode). All previous calibration data is
lost.
POLARIZATION
Contacting conductivity sensors only. Shows
information about the electrode polarization, the cable
capacitance and the time before the next
measurement.
SENSOR SIGNALS
Shows the current sensor signal information.
FACTORY CAL
Reserved for service technicians only.
DIAG MEAS
Shows diagnostic information about the current
measurement.
Error list
Errors may occur for various reasons. An error icon consists of an exclamation point within a circle.
When an error occurs, the error icon and the measurement screen flash alternately in the main
display. All outputs are held when specified in the controller menu. To view errors, push the
menu
key and select DIAGNOSTICS. Then select the device to view any problems associated with that
device.
A list of possible errors is shown in
.
English
19
Содержание Polymetron 9526
Страница 129: ...EMC A A A A EMC Canadian Radio Interference Causing Equipment Regulation IECS 003 A A FCC 15 A FCC 15 1 2 FCC 15 A 129...
Страница 130: ...1 2 3 4 5 2 DN8 DN6 DN6 2 IN OUT 100 s cm CO2 1 2 S cm KCI 10 S cm 130...
Страница 133: ...OUT 1 OUT 2 3 90 8 mm D8 D6 6mm 70 PTFE 4 OUT 5 D1 D2 2 2 20 L 60L 30 2 1 D1 2 D2 OUT 20L 60L 30 133...
Страница 134: ...3 0 35 2 mm2 AWG 22 14 105 4 7 1 131 4 1 5 9 2 6 10 3 7 4 8 1 6 4 6 2 1 4 134...
Страница 136: ...5 1 5 Back 2 6 Menu Settings Menu 3 HOME Main Measurement 7 4 ENTER SD 6 136...
Страница 206: ...EMC 206...
Страница 207: ...EMC A A IECS 003 A FCC 15 15 FCC 1 2 15 FCC 1 2 3 4 5 207...
Страница 208: ...2 DN8 DN6 DN6 2 100 CO2 1 2 10 KCl 208...
Страница 212: ...30 2 1 D1 2 D2 OUT 20 60 30 212...
Страница 213: ...3 0 35 2 2 AWG 22 14 105 C 4 7 1 209 213...
Страница 215: ...Startup 203 1 2 3 4 5 5 5 1 5 BACK 2 SD 6 MENU 3 HOME 7 4 ENTER SD 215...
Страница 216: ...6 6 1 7 2 8 3 SD 9 4 10 5 11 6 12 1 SD SD SD SD SD SD ENTER MENU Diagnostics Main Measurement 216...
Страница 245: ......