Table 4 Warning list for conductivity sensors (continued)
Warning
Description
Resolution
WRONG LINEAR TC
The user-defined linear temperature
compensation is out of range
The value must be between 0 and
4%/°C; 0 to 200 °C.
WRONG TC TABLE
The user-defined temperature
compensation table is out of range
The temperature is above or below
the temperature range defined by
the table.
WRNG USER CONC TABLE
The concentration measurement is
outside of the range of the user
table
Make sure that the user table is set
for the correct measurement range.
WRNG BLT-IN TEMP TABLE
The measured temperature is
outside of the range of the built-in
temperature compensation table
Make sure that the temperature
compensation is configured
correctly.
WRNG BLT-IN CONC TABLE
The concentration measurement is
outside of the range of the built-in
concentration table
Make sure that the concentration
measurement is configured for the
correct chemical and range.
Replacement parts and accessories
Refer to the replacement parts and accessories section of the controller documentation for controller
parts and accessories.
Note: Product and article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts and accessories
Description
Item no.
Kit of 3 protection caps for instrument front panel connectors
09126=A=8010
Kit of 2 adapters D6/8 to DN4/6
09126=A=8020
Kit of 2 black protection caps for sample input and output connectors
09126=A=8030
Power supply connector
350=500=004
Tool for disconnecting in/out sampling tubes
578=507=602
DN8 semi-rigid PTFE tubing (per meter)
590=060=080
DN8 semi-rigid PE tubing (per meter)
151400,22387
Cable for the 4-20 mA output (5 meters)
08319=A=0005
Cable for the 4-20 mA output (10 meters)
08319=A=0010
Cable for the 4-20 mA output (20 meters)
08319=A=0020
¾’’ NPT flow chamber in PP with fittings
09126=A=0100
Pt100 temperature simulator (0.1°C accuracy)
037=000=001
Cable for connecting the Pt100 temperature simulator
09125=A=8020
Yearly re-calibration in our factory
09526=A=1000
English
21
Содержание Polymetron 9526
Страница 129: ...EMC A A A A EMC Canadian Radio Interference Causing Equipment Regulation IECS 003 A A FCC 15 A FCC 15 1 2 FCC 15 A 129...
Страница 130: ...1 2 3 4 5 2 DN8 DN6 DN6 2 IN OUT 100 s cm CO2 1 2 S cm KCI 10 S cm 130...
Страница 133: ...OUT 1 OUT 2 3 90 8 mm D8 D6 6mm 70 PTFE 4 OUT 5 D1 D2 2 2 20 L 60L 30 2 1 D1 2 D2 OUT 20L 60L 30 133...
Страница 134: ...3 0 35 2 mm2 AWG 22 14 105 4 7 1 131 4 1 5 9 2 6 10 3 7 4 8 1 6 4 6 2 1 4 134...
Страница 136: ...5 1 5 Back 2 6 Menu Settings Menu 3 HOME Main Measurement 7 4 ENTER SD 6 136...
Страница 206: ...EMC 206...
Страница 207: ...EMC A A IECS 003 A FCC 15 15 FCC 1 2 15 FCC 1 2 3 4 5 207...
Страница 208: ...2 DN8 DN6 DN6 2 100 CO2 1 2 10 KCl 208...
Страница 212: ...30 2 1 D1 2 D2 OUT 20 60 30 212...
Страница 213: ...3 0 35 2 2 AWG 22 14 105 C 4 7 1 209 213...
Страница 215: ...Startup 203 1 2 3 4 5 5 5 1 5 BACK 2 SD 6 MENU 3 HOME 7 4 ENTER SD 215...
Страница 216: ...6 6 1 7 2 8 3 SD 9 4 10 5 11 6 12 1 SD SD SD SD SD SD ENTER MENU Diagnostics Main Measurement 216...
Страница 245: ......