K Calculation
Use this option to re-calculate the cell constant K value for the sensor under test.
1.
Push the
menu
key and select SENSOR SETUP>K CALCULATION.
Option
Description
K CALCULATION
This option is only valid if the last sensor calibration date is within one month of the
current date. The following parameters are required:
• SITE ID—The name of the site ID is limited to 10 characters in any combination of
letters, numbers, spaces or punctuation
• CONDUCTIVITY— Enter the measurement value from the controller under test
• TEMPERATURE— Enter the sample temperature from the controller under test
• CELL K VALUE— Enter the cell constant K value from the controller under test
• CALIBRATION SLOPE— Enter the slope value from the controller under test
The new K value is calculated and displayed and should be entered into the controller
under test.
Note: The calculation will fail if the new value is outside ±10% of the original value.
K CALCULATION
LOG
Lists all log files sorted by date and time. Use the arrow keys to select a log file and
push
enter
to view the calculation details.
RESET K CALC
LOG
Enter the factory pass code and select
YES
to delete the existing log file. Push
enter
to continue.
Maintenance
D A N G E R
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the document.
Cleaning the controller
D A N G E R
Always remove power from the controller before performing maintenance activities.
Note: Never use flammable or corrosive solvents to clean any part of the controller. Use of these solvents may
degrade the environmental protection of the unit and may void the warranty.
1.
Make sure the controller cover is securely closed.
2.
Wipe the controller exterior with a cloth dampened with water, or with a mixture of water and mild
detergent.
Clean the sensor
W A R N I N G
Chemical hazard. Always wear personal safety protection in accordance with the Material Safety Data Sheet for
the chemical that is used.
W A R N I N G
Personal injury hazard. Removal of a sensor from a pressurized vessel can be dangerous. Reduce the process
pressure to below 10 psi before removal. If this is not possible, use extreme caution. Refer to the documentation
supplied with the mounting hardware for more information.
Pre-requisite:
Prepare a mild soap solution with a non-abrasive dishwashing detergent that does not
contain lanolin. Lanolin leaves a film on the electrode surface that can degrade the sensor
performance.
Examine the sensor periodically for debris and deposits. Clean the sensor when there is a buildup of
deposits or when performance has degraded.
18
English
Содержание Polymetron 9526
Страница 129: ...EMC A A A A EMC Canadian Radio Interference Causing Equipment Regulation IECS 003 A A FCC 15 A FCC 15 1 2 FCC 15 A 129...
Страница 130: ...1 2 3 4 5 2 DN8 DN6 DN6 2 IN OUT 100 s cm CO2 1 2 S cm KCI 10 S cm 130...
Страница 133: ...OUT 1 OUT 2 3 90 8 mm D8 D6 6mm 70 PTFE 4 OUT 5 D1 D2 2 2 20 L 60L 30 2 1 D1 2 D2 OUT 20L 60L 30 133...
Страница 134: ...3 0 35 2 mm2 AWG 22 14 105 4 7 1 131 4 1 5 9 2 6 10 3 7 4 8 1 6 4 6 2 1 4 134...
Страница 136: ...5 1 5 Back 2 6 Menu Settings Menu 3 HOME Main Measurement 7 4 ENTER SD 6 136...
Страница 206: ...EMC 206...
Страница 207: ...EMC A A IECS 003 A FCC 15 15 FCC 1 2 15 FCC 1 2 3 4 5 207...
Страница 208: ...2 DN8 DN6 DN6 2 100 CO2 1 2 10 KCl 208...
Страница 212: ...30 2 1 D1 2 D2 OUT 20 60 30 212...
Страница 213: ...3 0 35 2 2 AWG 22 14 105 C 4 7 1 209 213...
Страница 215: ...Startup 203 1 2 3 4 5 5 5 1 5 BACK 2 SD 6 MENU 3 HOME 7 4 ENTER SD 215...
Страница 216: ...6 6 1 7 2 8 3 SD 9 4 10 5 11 6 12 1 SD SD SD SD SD SD ENTER MENU Diagnostics Main Measurement 216...
Страница 245: ......