Table 3 Error list for conductivity sensors
Error
Description
Resolution
ADC FAILURE
The analog to digital conversion
failed
Make sure that the sensor module is
fully inserted into the controller
connector. Replace the sensor
module.
SENSOR MISSING
The sensor is missing or
disconnected
Examine the wiring and connections
for the sensor and for the module.
Make sure that the terminal block is
fully inserted into the module.
SENS OUT RANGE
The sensor signal is outside of the
accepted limits (2 S/cm)
Make sure that the display format is
set for the correct measurement
range.
Warning list
A warning icon consists of an exclamation point within a triangle. Warning icons appear on the right
of the main display below the measurement value. A warning does not affect the operation of menus,
relays and outputs. To view warnings, push the
menu
key and select DIAGNOSTICS. Then select
the device to view any problems associated with that device. The warning icon will no longer be
displayed once the problem has been corrected or acknowledged.
A list of possible warnings is shown in
Table 4 Warning list for conductivity sensors
Warning
Description
Resolution
MEAS TOO HIGH
The measured value is > 2 S/cm,
1,000,000 ppm, 200% or 20,000 ppt
Make sure that the display format is
set for the correct measurement
range
MEAS TOO LOW
The measured value is < 0 μS/cm,
0 ppm, 0% or 0 ppt
Make sure that the sensor is
configured for the correct cell
constant.
ZERO TOO HIGH
The zero calibration value is too
high
Make sure that the sensor is held in
air during zero calibration and is not
located near radio frequency or
electromagnetic interference. Make
sure that the cable is shielded by
metal conduit.
ZERO TOO LOW
The zero calibration value is too low
TEMP TOO HIGH
The measured temperature is >
200 °C
Make sure that the sensor is
configured for the correct
temperature element.
TEMP TOO LOW
The measured temperature is <
-20 °C
CAL OVERDUE
The Cal Reminder time has expired
Calibrate the sensor.
NOT CALIBRATED
The sensor has not been calibrated
Calibrate the sensor.
REPLACE SENSOR
The sensor has been in operation >
365 days
Calibrate the sensor with a
reference solution and reset the
sensor days. Refer to
on page 19. If the calibration fails,
call technical support.
CAL IN PROGRESS
A calibration was started but not
completed
Return to calibration.
OUTPUTS ON HOLD
During calibration, the outputs were
set to hold for a selected time.
The outputs will become active after
the selected time period.
20
English
Содержание Polymetron 9526
Страница 129: ...EMC A A A A EMC Canadian Radio Interference Causing Equipment Regulation IECS 003 A A FCC 15 A FCC 15 1 2 FCC 15 A 129...
Страница 130: ...1 2 3 4 5 2 DN8 DN6 DN6 2 IN OUT 100 s cm CO2 1 2 S cm KCI 10 S cm 130...
Страница 133: ...OUT 1 OUT 2 3 90 8 mm D8 D6 6mm 70 PTFE 4 OUT 5 D1 D2 2 2 20 L 60L 30 2 1 D1 2 D2 OUT 20L 60L 30 133...
Страница 134: ...3 0 35 2 mm2 AWG 22 14 105 4 7 1 131 4 1 5 9 2 6 10 3 7 4 8 1 6 4 6 2 1 4 134...
Страница 136: ...5 1 5 Back 2 6 Menu Settings Menu 3 HOME Main Measurement 7 4 ENTER SD 6 136...
Страница 206: ...EMC 206...
Страница 207: ...EMC A A IECS 003 A FCC 15 15 FCC 1 2 15 FCC 1 2 3 4 5 207...
Страница 208: ...2 DN8 DN6 DN6 2 100 CO2 1 2 10 KCl 208...
Страница 212: ...30 2 1 D1 2 D2 OUT 20 60 30 212...
Страница 213: ...3 0 35 2 2 AWG 22 14 105 C 4 7 1 209 213...
Страница 215: ...Startup 203 1 2 3 4 5 5 5 1 5 BACK 2 SD 6 MENU 3 HOME 7 4 ENTER SD 215...
Страница 216: ...6 6 1 7 2 8 3 SD 9 4 10 5 11 6 12 1 SD SD SD SD SD SD ENTER MENU Diagnostics Main Measurement 216...
Страница 245: ......