EMR Setup
(continued)
Setup
1. Tap
Settings
, located at the top right of the Controller screen.
2. Tap
Ports
on the Settings menu.
3. In the Port Selection panel, tap
EMR
. Refer to Figure IS-25.
4. In the Port Settings panel, select the device according to the list below.
Devices
Visutron 900
Touch
- Sends Visutron 900
Touch
final refraction data only (including subjective re-
fraction data and binocular test results).
Visutron 900
Touch
Exp
- Sends Visutron 900
Touch
final refraction data (including subjective re-
fraction data and binocular test results)
and
pre-test data.
AutoPhor
- Sends Visutron 900
Touch
final refraction data only.
AutoPhor Exp
- Sends Visutron 900
Touch
final refraction data
and
pre-test data.
Vis900
- Sends data in original Visutron 900+ format, for Visutron 900 units
only
.
5. Connect the computer to the EMR port on the Central Unit with a cable or Bluetooth dongle.
a. Refer to the
Connecting with Cables
section if connecting with a cable.
b. Refer to the
Connecting with Bluetooth Dongles
section if connecting with Bluetooth.
6. Set the serial communication parameters to match the settings of the computer serial port.
7. Tap
Apply
.
Figure IS-25, Port Settings Screen
Instrument Setup
(continued)
20
16243-101 Rev. G