PC Interface
LT51 User Manual
75
READDATA1
READDATA2
READDATA3
Syntax: READDATA1
<0xB1><xx><xx><xx><xx>
Syntax: READDATA2
<0xB2><xx><xx><xx><xx>
Syntax: READDATA3
<0xB3><xx><xx><xx><xx>
LT51 returns:
<D0><D1><D2><D3>
The LT51 will return 4 bytes (signed 32-bit value with MSB D3, LSB D0)
representing the data for:
READDATA1
– the
currently
selected terminals
READDATA2
– the AG terminals
READDATA3
– the BG terminals
depending on the following current measurement mode:
Command
Mode
Data
READDATA1
READDATA2
READDATA3
R & C
R = (D3 * 256) + D2
C = (D1 * 256) + D0 nF
If (R < 500)
R & C are in series
Else
R & C are in parallel
VOLTAGE
DCV = [((D1 * 256) + D0) / 100] V, NB: signed
ACV = [((D3 * 256) + D2) / 100] V
READDATA1
MAGZ
32-bit Data
READDATA1
Others
(32-bit Data / 100) dB / dBm / none
READDATA2
Others
32-bit Data (Hz), NB: = Frequency at RTX
READDATA3
Others
(32-bit Data / 100) dB / dBm / dHz
Value at Marker or Distortion if set