SECTION11 ERROR AND FAULT CODES
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Page 11-3
200
65
INTERNAL FAILURE
[R]
RFBUS_VOLTS failure [POST check]
200
66
INTERNAL FAILURE
[R]
POST AUDIO Setup [POST check]
200
67
INTERNAL FAILURE
[F]
PSU_STATUS failure [POST check]
200
68
INTERNAL FAILURE
[F]
Mains input failure [POST check]
200
69
INTERNAL FAILURE
[F]
OVERDOSE failure [POST check]
200
70
INTERNAL FAILURE
[F]
PWM low failure [POST check]
200
71
INTERNAL FAILURE
[F]
ID CAL circuit failure [POST check]
200
72
INTERNAL FAILURE
[F]
BUSLIM failure [POST check]
200
73
INTERNAL FAILURE
[F]
CURRLIM failure [POST check]
200
74
INTERNAL FAILURE
[F]
VOLTLIM failure [POST check]
200
75
INTERNAL FAILURE
[F]
SPRF_ACTIVE failure [POST check]
200
76
INTERNAL FAILURE
[F]
VCC_ANA_DIV_2 failure [POST check]
200
77
INTERNAL FAILURE
[F]
12V failure [POST check]
200
78
INTERNAL FAILURE
[F]
0V failure [POST check]
200
80
INTERNAL FAILURE
[F]
RF_VOLTAGE [POST check]
200
81
INTERNAL FAILURE
[F]
RF_CURRENT failure [POST check]
200
82
INTERNAL FAILURE
[F]
ANALOGUE_REF failure
200
83
INTERNAL FAILURE
[F]
VCC_ANA_DIV_2 failure
200
84
INTERNAL FAILURE
[F]
12V failure
200
85
INTERNAL FAILURE
[F]
0V failure
200
86
INTERNAL FAILURE
[F]
RF_DET stuck high failure
200
87
INTERNAL FAILURE
[F]
PSU_STATUS failure
200
88
INTERNAL FAILURE
[F]
Mains input failure
200
89
INTERNAL FAILURE
[F]
BUSLIM failure
200
90
INTERNAL FAILURE
[F]
CURRLIM_SET failure
200
91
INTERNAL FAILURE
[F]
SP_PSU_OR_PK_RELAY failure
200
92
INTERNAL FAILURE
[F]
Electrode ID Failure
300
10
THERMAL SHUTDOWN
[R]
Internal overheating
300
11
INTERNAL FAILURE
[F]
Excess RF input voltage error (RFBUS > set)
300
12-13
Not used
---
-----
300
14
OUTPUT SHORTED
[R]
Accessory error : excessive RF output current
300
15
INTERNAL ERROR
[F]
Impedance V,I feedback circuit error
300
16
INTERNAL ERROR
[R]
SP board not supplying RF energy
300 17-19
Reserved
300
20
INVALID ELECTRODE
[SOFT] Unsupported electrode type
300
21
SP OUTPUT ERROR
[R]
Persistent over voltage or current error
400
10
FOOTPEDAL STUCK
[S]
Footswitch BLUE pedal stuck
400
11
FOOTPEDAL STUCK
[S]
Footswitch YELLOW pedal stuck
400
12
FOOTPEDAL STUCK
[S]
Footswitch MODE pedal stuck
400
14
INTERNAL FAILURE
[F]
Electrode identification circuit fault
400
15
BUTTON STUCK
[R]
Front panel : CUT (left) UP button stuck
400
16
BUTTON STUCK
[R]
Front panel : CUT (left) DOWN button stuck
400
17
BUTTON STUCK
[R]
Front panel : blue UP button stuck
400
18
BUTTON STUCK
[R]
Front panel : blue DOWN button stuck