SECTION 8 AFTER SURGERY
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Page 8-2
8D. Cleaning and Sterilizing the PK Connector Cable
Prepare the PK connector cable for steam sterilization according to the following cleaning
procedure:
a)
Remove all gross matter (blood, mucus, tissue etc.) by wiping the PK Connector
Cable with a cloth or gauze pad and a mild cleaning solution or detergent capable of
removing organic deposits.
IMPORTANT
To avoid damage to the cable and connectors do not immerse in
reprocessing solutions or use abrasive cleaning agents.
b)
Rinse thoroughly in running water.
c)
Remove residual cleaning agents with a water damp cloth.
d)
Dry the device thoroughly before sterilizing.
Sterilize the connector cable according to the sterilization procedure described in the
Connector Cable Instructions For Use.
IMPORTANT
The connector cable supplied as part of the Gyrus ACMI PlasmaKinetic
SuperPulse System is intended for 20 re-use cycles only.
WARNING
Do not attempt to re-use PlasmaKinetic
Instruments. Heat or chemical
sterilization may render the instrument mechanically or electrically
unsafe.
WARNING
Exceeding the recommended number of uses may result in electrical or
mechanical failure during use or difficulty when attaching or detaching
the instrument to or from the PK connector cable.
SECTION 8
AFTER SURGERY
Содержание PlasmaKinetic SuperPulse
Страница 2: ......