SECTION 14 EMC TABLES
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The SuperPulse Generator is intended for use in the electromagnetic environment specified below. The
customer or the user of the SuperPulse Generator should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV Contact
±8 kV Air
±6 kV Contact
±8 kV Air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical Fast
Transient / Burst
IEC 61000-4-4
±2 kV for Power
Supply Lines
±1 kV for Input /
Output Lines
±2 kV for Power
Supply Lines
±1 kV for Input /
Output Lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV Differential
Mode
±2 kV Common
Mode
±1 kV Differential
Mode
±2 kV Common
Mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Input
Lines
IEC 61000-4-11
<5 %
U
T
(>95 % Dip in
U
T
) for
0.5 Cycle
40 %
U
T
(60 % Dip in
U
T
) for
5 Cycle
70 %
U
T
(30 % Dip in
U
T
) for
25 Cycle
<5 %
U
T
(>95 % Dip in
U
T
) for
5 sec
<5 %
U
T
(>95 % Dip in
U
T
)
for 0, 5 Cycle
40 %
U
T
(60 % Dip in
U
T
) for
5 Cycle
70 %
U
T
(30 % Dip in
U
T
) for
25 Cycle
<5 %
U
T
(>95 % Dip in
U
T
)
for 5 sec
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
SuperPulse Generator requires
continued operation during power
mains interruptions, it is
recommended that the SuperPulse
Generator be powered from an
uninterruptible power supply or a
battery.
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Page 14-2
SECTION 14
EMC TABLES