SECTION 2 PATIENT AND OPERATING ROOM SAFETY
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Page 2-3
Flammable anesthetics or oxidizing gases such as nitrous oxide
(N
2
O) and oxygen enriched atmospheres.
Some materials, for example cotton, wool and gauze, when
saturated with oxygen may be ignited by sparks produced in normal
use of electrosurgical equipment.
The sparking and heating associated with electrosurgery can provide an ignition source.
Observe fire precautions at all times.
WARNING
Fire/Explosion Hazard: Verify that all oxygen circuit connections are leak free
before and during use of electrosurgery. When using electrosurgery in the same
room with any of the above substances or gases, prevent their accumulation or
pooling under surgical drapes, or within the area where electrosurgery is being
performed.
2D. Before
Surgery
Active Accessories
WARNING
Electric Shock Hazard: Do not connect wet accessories to the generator.
WARNING
Electric Shock Hazard: Ensure that all accessories are correctly connected and
that no metal is exposed.
WARNING
Do not attempt to re-use instruments labeled for Single Use Only. Heat or
chemical Sterilization may render the instrument mechanically or electrically
unsafe
CAUTION
Read the instructions, warnings and cautions provided with the instrument
accessories before using.
CAUTION
Accessories labeled as re-usable must only be processed according to the
recommended procedure and, where appropriate, recycled the specified number
of times.
CAUTION
Use default power levels to test an accessory.
CAUTION
Always inspect the system accessories for damage prior to use. In particular,
check the cables of any re-usable accessory for possible insulation damage.
CAUTION
Use only Gyrus ACMI approved accessories supplied for use with this product.
Product damage or accessory failure may otherwise result during use.
SuperPulse Generator
WARNING
Electric Shock Hazard; Connect the generator power cord to a properly
grounded receptacle. Do not use power plug adapters.
WARNING
Fire Hazard; Do not use extension cords.
CAUTION
Provide as much distance as possible between the generator and other
electronic equipment (such as monitors) as an activated generator may cause
interference with them.
CAUTION
Non-function of the generator may cause interruption of surgery. Ensure that all
installation procedures are followed and that all connectors are correctly inserted
before use. A backup generator should be available for use.
SECTION 2
PATIENT AND OPERATING ROOM SAFETY