SECTION 10 PERFORMANCE SPECIFICATIONS
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Page 10-2
Max Voltage
PK1
360 Volts Peak
PK2
434
Volts
Peak
PK3
480
Volts
Peak
T1
434
Volts
Peak
T2
480
Volts
Peak
DES
170
Volts
Peak
VP1
103
Volts
Peak
VP2
141
Volts
Peak
VP3
170
Volts
Peak
MR2
480
Volts
Peak
CAUTION
The following load curves apply to the fundamental power delivery capability of
the generator alone. They do not imply a given power output for any given
instrument and connector cable configuration when used with the generator. Each
accessory will self-impose an upper set power limit for the generator.
CAUTION
Some electrode settings will not allow full power to be administered.
Power Linearity 230V 50 Hz a.c
.
0
50
100
150
200
250
0
50
100
150
200
250
Measured Power
Set Power
Linearity (PK3 into 150)
SECTION 10