SECTION 5 BEFORE SURGERY
Gyrus ACMI PlasmaKinetic SuperPulse Generator
USER MANUAL
Part Number: 144012-MB
Page 5-1
This section describes how to set up the SuperPulse Generator before surgery. Prior to using the
system, you should ensure that the following associated equipment has been prepared for use:
•
An appropriate sterilized reusable Connector Cable.
•
An appropriate Gyrus ACMI PK instrument(s) for the procedure to be performed.
5A. Power up the Generator
Install the Generator
•
Place the generator on a table, cart, racking system or other stable platform that can
be positioned as close as possible to the operative site during use.
•
Ensure that the generator is standing on a flat, firm surface and that it is not at risk of
being accidentally dislodged during the course of operating theatre activity.
•
Provide at least four inches of space from the rear of the generator. Never cover the
generator or stack other equipment on top of it. It is normal for the generator to
become warm during use so ensure adequate ventilation.
Connect the Generator
Plug the supplied AC power cord into the receptacle on the rear of the generator and then
connect the other end of the cord directly to an AC power point. Avoid the use of extension
cords or multiple plug adaptors. Wherever possible, avoid trailing leads and neatly store
excess power cord.
Connect the Footswitch
Connect the footswitch to the receptacle at the rear of the generator by orienting the lug on
the plug with the groove in the receptacle and pushing the plug into place. Secure the plug
by screwing down the locking ring in a clockwise direction.
Switch the Generator On
With the rear panel switch on,.
Press the Standby / On button and verify the completion of the system initialization. The
system is now ready for use with any connected instrument.
5B. Select and Connect the Connector Cable
Depending on which instruments are to be used select either a 5 or 3-way PK connector
cable.
Introduce the sterile connector cable to the sterile instrument table or trolley according to
the sterile handling practices at your facility.
The cable and connectors should be inspected for any processing damage.
IMPORTANT
If any of the connector pins are bent or if the cable shows any signs of
crush damage, cracking or distortion, it must be discarded. The
generator end of the connector cable should then be passed to a non-
sterile operative for connection to the generator. Ensure that sufficient
length of cable is retained for connection to the instrument, that
sufficient slack is provided so that operation is not impeded, and that
sufficient slack is also allowed for attachment to the surgical drapes.
5C. Attach Gyrus ACMI PK Instrument to PK Connector Cables
Connect the Gyrus ACMI PK instrument to the PK Connector Cable by aligning the two
connector halves and pushing together. Once connection is made the generator display
will change to the default settings appropriate to the instrument being used.
SECTION 5
BEFORE SURGERY