OPERATION
197
4.
Use the Selector knob to change
Total Unit from the OFF setting to
the number of parallel units.
U
s
Total Unit
2
5.
Press the selector knob or Enter to
confirm.
ENTER
or
Parallel Mode
6.
To change the type of mode, use
the Selector knob to edit Group
Mode.
7.
Choose Para to operate the units as a single
large load module, or choose Sync to
synchronize the load settings across each
parallel unit.
U
s
Group Mode
Para
Display Mode
8.
Use the selector knob to change
the display settings on the local
load modules.
9.
Choose from V, I / V, W / I,W or S.
U
s
Display Mode
V,I
When Para Unit is active, an indicator
will be displayed on the screen. The
indicator depends on the Group Mode.
P will be displayed for Para Mode and
S will be displayed for Sync Mode.
CCDH
CH1P
conf
80V
CHXP
CHXS
Para Mode
Sync Mode
Содержание PEL-2000B Series
Страница 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 5: ...Table of Contents 5 INDEX 312...
Страница 13: ...GETTING STARTED 13 4 panel setups USB flash drive support Interface USB RS 232C RS485 LAN GPIB optional...
Страница 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Страница 307: ...APPENDIX 307 Input Resistance Load OFF 500k Typical Temperature Coefficient 100ppm Weight Approx 3 8kg...
Страница 308: ...PEL 2000B Series User Manual 308 Dimensions PEL 2002B 272 177 532 3 195 3 554 1 318 4 533 3 583 3...
Страница 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Страница 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Страница 317: ...INDEX 317 Wiring procedure 46...