PEL-2000B Series User Manual
126
Panel Operation
1.
Press the F2 (Range) key repeatedly
until High or Low range is selected.
Range
High
Range
Low
F1
F5
~
F2
The range will be reflected in both the
bottom menu system and the Current
Operation Channel Status panel.
CCDH
main
80V
CH1P
CC Dynamic Low
CC Dynamic High
CC Static Low
CC Static High
CCDL
CCDH
CCL
CCH
Note
Changing the range will only affect the current
(active) channel. Other channels will not be affected
by any changes.
Not all load modules support dual ranges. If only one
range is supported, it is usually high range.
Select CC Dynamic Mode
Background
Constant current mode can be set to dynamic or
static mode. Dynamic mode is used to
automatically set varying load rates.
Ensure the menu is in CC Mode. See page 125.
Configure
Configure
Mode
CC
Dynamic
Range
High
Panel Operation
1.
Press the F3 key until Dynamic
Range mode is selected.
Dynamic
Static
F1
F5
~
F3
Содержание PEL-2000B Series
Страница 1: ...Programmable DC Electronic Load PEL 2000B Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER...
Страница 5: ...Table of Contents 5 INDEX 312...
Страница 13: ...GETTING STARTED 13 4 panel setups USB flash drive support Interface USB RS 232C RS485 LAN GPIB optional...
Страница 113: ...TUTORIALS 113 14 Sound Turn the sound on off for the mainframe IU Page 202...
Страница 307: ...APPENDIX 307 Input Resistance Load OFF 500k Typical Temperature Coefficient 100ppm Weight Approx 3 8kg...
Страница 308: ...PEL 2000B Series User Manual 308 Dimensions PEL 2002B 272 177 532 3 195 3 554 1 318 4 533 3 583 3...
Страница 309: ...APPENDIX 309 PEL 2004B 435 4 177 532 3 195 3 481 8 554 1 533 3 583 3...
Страница 310: ...PEL 2000B Series User Manual 310 PEL 2020B PEL 2030B PEL 2040B PEL 2041B 81 6 171 2 475 446 8...
Страница 317: ...INDEX 317 Wiring procedure 46...